BraggMONITOR SI User Interface
FS22 SI
A4249-8.0
HBM: public
99
ventional method (ranges OFF) that uses a global thresh
old and the Smart Peak Detection (ranges ON) that uses
the threshold individually inside each sensor range.
Ranges
Using the “ranges on/off" button (number
3
the user may select which FBG sensor detection method
will be used.
When ranges are ON (
will display the range associated with each sensor that is
listed in the Configuration tab and the respective thresh
old used for the FBG peak computation. Even when mul
tiple peaks occur inside one range, only one FBG sensor
will be detected.
When the ranges are OFF (
graph will contain the optical spectrum of the selected
channel and an horizontal line representing the threshold
level considered for the detection of the FBG sensors.
On the right side of the spectral view tab, below the opti
cal channel selector, one can find info regarding the
detected sensors (number
12
in
).
When the conventional method for FBG sensor detection
is selected only the “Wavelengths" table is displayed
(
). In that table it is possible to observe the
wavelengths of the detected sensor peaks (WL (nm)) and
the corresponding powers (PWR (dBm)). Below the
“Wavelengths" table a numeric indicator shows the num
ber of detected sensors on the selected channel (num
ber
11
in
).
Conventional
Measurement
Method
Summary of Contents for FS22 SI
Page 1: ...User Manual English FS22 SI Industrial BraggMETER SI...
Page 22: ...Operation 22 A4249 8 0 HBM public FS22 SI Fig 3 6...
Page 92: ...BraggMONITOR SI User Interface 92 A4249 8 0 HBM public FS22 SI Fig 7 17 Fig 7 18...
Page 93: ...BraggMONITOR SI User Interface FS22 SI A4249 8 0 HBM public 93 Fig 7 19 Fig 7 20...