Operation
18
A4249-8.0
HBM: public
FS22 SI
Important
The interrogator default network configuration is
“10.0.0.10:255.0.0.0:0.0.0.0".
3.3
Switching On
Pressing the ON/OFF button will start the interrogator.
The STATUS LED will start blinking at 2 Hz. After
approximately 30 seconds it will start blinking at 1 Hz.
This means that the interrogator is already on and
responsive, but the optoelectronic module is still warming
up. After approximately 90 seconds it should stay on per
manently. This means that the interrogator is able to
measure.
Status LED
Blinking 2 Hz
Status LED
Blinking 1 Hz
Status LED
On
Status LED
Blinking 2 Hz
Status LED
Blinking 1 Hz
Status LED
Blinking 5 Hz
(initialization issues)
Correct
Wrong
ON(0s)
~10s
>30s
<90s
?s
Status LED
Blinking 5 Hz
(measurement issues)
Information
If the interrogator does not start correctly, the STATUS
LED will blink faster. If this happens please contact HBM
FiberSensing technical support.