Operation
24
A4249-8.0
HBM: public
FS22 SI
3.5.2
Resetting the IP Address of the Interrogator
If for any chance there is the need to physically change
the IP address of the interrogator, there is a reset proce
dure that consists on pressing the ON/OFF button for
more than 10 seconds.
When the button is being pressed for 10 seconds the
POWER LED blinks 3 times and goes OFF. At this
moment the ON/OFF button can be released and the
reset of IP address will happen.
POWER LED
blinking at 1 Hz
POWER LED
ON
POWER LED
blinks 3 times
and goes OFF
Fig. 3.8
This procedure resets the interrogator changing its IP
address to the default and its measuring settings to the
last stored.
Information
The interrogator default network properties are
“10.0.0.10:255.0.0.0:0.0.0.0".