Remote Communications
FS22 SI
A4249-8.0
HBM: public
71
:
ACQUisition:SAVE:CONTinuous:STARt
Starts continuous absolute wavelength peak saving on the
instrument internal memory. Saved data is stored on a sepa
rate directory for each day. The directory name is the date in
the format «YYYYMMDD»
Note: Saving with different operating methods may be saved
on the same data set, if these two types of data recording
have been selected on the same day.
Note: Scheduled acquisition may be saved on the same data
set as continuous acquisition if these two types of data
recording have been selected on the same day (
see section
“Scheduled Acquisition“ on page 71
Example:
Command
:ACQU:SAVE:CONT:STAR
Answer
Port 3500 » :ACK
Scheduled Acquisition
:ACQUisition:SCHEdule:STARt
Starts scheduled acquisition with the defined period and
number of samples. Data is stored on the instrument internal
memory on a separate directory for each day. The directory
name is the date in the format «YYYYMMDD».
Note: Scheduled acquisition may be saved on the same data
set as continuous acquisition if these two types of data
recording have been selected on the same day (
see section
“Continuous Acquisition“ on page 64
The interrogator does not resume scheduled acquisition if
power fails during sleep.
Example:
Command
:ACQU:SCHE:STAR
Answer
:ACK
Summary of Contents for FS22 SI
Page 1: ...User Manual English FS22 SI Industrial BraggMETER SI...
Page 22: ...Operation 22 A4249 8 0 HBM public FS22 SI Fig 3 6...
Page 92: ...BraggMONITOR SI User Interface 92 A4249 8 0 HBM public FS22 SI Fig 7 17 Fig 7 18...
Page 93: ...BraggMONITOR SI User Interface FS22 SI A4249 8 0 HBM public 93 Fig 7 19 Fig 7 20...