![HBK U9C Mounting Instructions Download Page 25](http://html1.mh-extra.com/html/hbk/u9c/u9c_mounting-instructions_2111685025.webp)
23
U9C
ELECTRICAL CONNECTION
8.2.2
Connecting the device
The device is connected via a M12 plug on the front of the inline amplifier. The assign
ment is set out in the following table. The supply voltage must be within the specified
range (19 ... 30 V).
Pin
Version VA 1
(voltage output)
Version VA 2
(current output)
KAB168
connection cable
wire assignment
1
Supply voltage 0 V (GND)
white
2
Not in use
brown
3
Control input zero setting
green
4
Not in use
yellow
5
Output signal
0 ... 10 V
Output signal
4 ... 20 mA
gray
6
Output signal
0 V
Not in use
pink
7
Not in use
blue
8
Voltage 19 ... +30 V
red
The cable connecting the inline amplifier to the next link in the measurement chain must
not exceed 30 m in length.
8.2.3
Operating the inline amplifier/zeroing the measurement chain
The measurement starts as soon as the sensor is connected to a supply voltage and the
output of the amplifier is connected to the next link in the measurement chain.
If you apply a voltage > 10 V to the “Zero” input, a one-time zeroing is performed. After
this zeroing, the device continues to measure, even if you apply a voltage above 10 V to
the input.
To trigger a new zeroing operation, the input must first be set to 0 V and then be reset by
applying a voltage of over 10 V.
Notice
Please note that you can zero the measurement chain with any force applied. If an initial
load is already acting on the force transducer, it is essential to consider it, otherwise the
force transducer may be overloaded.
The zero point is not permanently stored in the device. If you have disconnected the mea
surement chain from the supply voltage, we recommend zeroing again.
Summary of Contents for U9C
Page 3: ...U9C ENGLISH DEUTSCH FRAN AIS ITALIANO Mounting Instructions...
Page 37: ...U9C ENGLISH DEUTSCH FRAN AIS ITALIANO Montageanleitung...
Page 69: ...U9C ENGLISH DEUTSCH FRAN AIS ITALIANO Notice de montage...
Page 105: ...U9C ENGLISH DEUTSCH FRAN AIS ITALIANO Istruzioni per il montaggio...
Page 107: ...3 U9C SOMMARIO 9 Dimensioni 28 10 Dati tecnici 33 10 1 Versioni e no ordine 37...
Page 143: ...39 U9C DATI TECNICI...
Page 144: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A03815 03 YI0 02 7 0111 0012...