MECHANICAL INSTALLATION
16
5
MECHANICAL INSTALLATION
Conditions at the installation site
S
Protect the device from direct contact with water.
S
Protect the device from moisture and weather such as rain or snow. The protection
class of the device is IP65 (DIN EN 60529).
S
Do not expose the device to direct sunlight.
Mounting position
The device can be mounted in any position.
Installation
Mount the device with two bolts with an outside diameter of less than 4.4mm. Remove
the cover of the housing to reach the mounting holes.
Please also see section 6.2 starting on page 19, which explains correct connection via
one of the PG glands.
After cable installation, tighten the screws of the housing cover to a torque of approx.
1 Nm to ensure the stated IP rating and the best possible EMC protection.