Troubleshooting
11
Troubleshooting
Fault
Possible cause
Remedy
Mains cable damaged
damage caused by third party,
transport damage
obtain original spare mains
cable from manufacturer or
supplier
No operating functions; all
LEDs dark
mains cable not plugged in
plug in mains cable
socket dead
check socket/fuse
mains cable
damaged/interrupted
replace mains cable
fault of electronics
return unit to supplier or
manufacturer
No ultrasonic operation;
LED ultrasound dark
turning knob for ultrasonic
operation in „0“ position
switch on ultrasonic
operation by turning the knob
key
►
■
(ultrasound) not
pressed
press key
►
■
fault of electronics
return unit to supplier or
manufacturer
No ultrasonic operation;
timer LEDs flash
alternatingly (“running
light”) = fault indication
ultrasound
fault of electronics
switch unit off and on; if fault
is indicated again return unit
to supplier or manufacturer
Unsatisfactory cleaning
results
no or unsuitable cleaning
medium used
use suitable cleaning
medium
cleaning period too short
repeat cleaning interval
ultrasonic energy heats up
liquid (physical process)
switch on ultrasound for
short period only