HAUBER-Elektronik GmbH
9
11
Technical data
11.1
General data
Each sensor has one of the listed measuring and frequency ranges. Further
ranges on request.
Please indicate the measuring and frequency range in your request.
Measuring range:
0-8 mm/s rms (only for frequency range > 10 Hz)
0-10 mm/s rms
0-16 mm/s rms
0-20 mm/s rms
0-25 mm/s rms
0-32 mm/s rms
0-50 mm/s rms
0-64 mm/s rms
0-128 mm/s rms
Measuring accuracy:
± 10% (as per DIN ISO 2954)
Transverse sensitivity:
< 5%
Frequency range:
10 Hz
–1,000 Hz (standard)
1 Hz
–1000 Hz
Calibration point:
159.2 Hz and 90% amplitude of measuring range
Bearing status parameter:
Weighted crest factor.
Crest factor
𝑘
𝑠
=
|𝑋|
𝑚𝑎𝑥
𝑋
𝑒𝑓𝑓
Weight / categorised by vibration intensity diagram from DIN
ISO 13373-3
Maximum acceleration:
±16.5 g
Lifetime:
10 years
MTTF value:
%placeholder%
Tab. 1: General data