Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic environment
–
guidance
Voltage dips,
Short
interruptions
and
voltage
variations
on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95%dip in U
T
)
For 0.5 cycle
40% U
T
(60%dip in U
T
)
For 5 cycles
70% U
T
(30%dip in U
T
)
For 25 cycles
<5% U
T
(>95%dip in U
T
)
<5% U
T
(>95%dip in U
T
)
For 0.5 cycle
40% U
T
(60%dip in U
T
)
For 5 cycles
70% U
T
(30%dip in U
T
)
For 25 cycles
<5% U
T
(>95%dip in U
T
)
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the TAURUS C1 requires
continued operation during power mains
interruptions, it is recommended that the
TA U R U S C 1 b e p o w e r e d f r o m a n
uninterruptible power supply or a battery.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical Commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150kHz to 80MHz
Outside ISM bands
3V/m
80MHz to 2.5GHz
3 Vrms
3 Vrms
Portable and mobile RF communications
dental chair unit should be used no closer
to any part of the TAURUS C1, including
cable, than the recommended separation
distance calculated from the equation
applications to the Frequency of the
transmitter.
Recommended separation distance
d=
1.2√P
d=
1.2√P
80MHz to 800MHz
d=
2.3√P
800MHz to 2.5GHz
Where P is the maximum output power
rating of the transmitter in watts(W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
¹
should be less than the compliance
level in each frequency range.
²
Interface may occur in the vicinity of dental
chair unit marked with the following symbol: