Medical electrical dental chair unit needs special precautions regarding EMC and needs to be installed
and Put into service according to the EMC information provided in this manual.
Portable and mobile RF communications dental chair unit can affect medical electrical dental chair
unit. The dental chair unit or system should not be used adjacent to or stacked with other dental
chair unit. If adjacent or stacked use is necessary, the dental chair unit or system should be observed
to verify normal operation in the configuration in which it will be used.
Guidance and manufacture`s declaration
–
electromagnetic emissions
The TAURUS C1 is intended for use in the electromagnetic environment specified below. The customer or the user
of the TAURUS C1 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TAURUS C1 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic.
RF emissions
Class B
CISPR 11
The TAURUS C1 is suitable for use in all establishments,
including domestic establishment and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacture`s declaration
–
electromagnetic immunity
The TAURUS C1 is intended for use in the electromagnetic environment specified below. The customer or the user
of the TAURUS C1 should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV
contact
±8kV
air
±6kV
VCP
±6kV
contact, VCP
±8kV
air
±6kV
VCP
Floors should be wood, concrete or ceramic
file. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
Supply lines
±1kV
for input/output
line
±2kV for
power
Supply lines
N.A.
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC61000-4-5
±1kV
differential mode
±2kV
common mode
±1kV
differential mode
±2kV
common mode
Mains power quality should be that of a
typical commercial or hospital environment.
12. Electromagnetic Compatibility (EMC)