Solution 2:
Move the drum to its home position. To do this, run
FlexColor and select
Maintenance > Feeder
to open the
Feeder
window. Then set the
Drum position
drop-down list
to “home”. Try scanning again.
Solution 3:
Make sure that the magazine is pressed fully into the drawer
hinge. See also “Loading Slides” on page 18.
Solution 4:
Check that light is not shining directly into the scanner. The
Feeder uses an optical system to monitor the position of the
original holder; this system can be confused if outside light is
shining directly into the scanner.
Solution 5:
Check the alignment of the Feeder to make sure the slide
original holder is centered. See “Aligning the Feeder” on page
30.