SERVICING
10 SERVICING
10.1 PRINCIPLES
Operators of care beds are obliged according to MPBetreibV (Operators of Medical Products
Ordinance) §4 to guarantee the safe condition of the medical product over their entire service
life.
The test according to the regulation EN 62353 contains the following minimum requirements:
• Visual check
• Measurement of leakage resistance
• Measurement of leakage current
• Functional test
• Overall evaluation
The service life of the care bed depends essentially upon the handling and servicing.
To guarantee safe operation, a visual and functional test including an electrical test must be
carried out at least once a year. For this purpose, proceed according to the technical safety
checklist as per regulation EN 62353 in section
10.2
If you have any doubts about the safety or functionality of the bed or
even a part of the bed as a result of the work performed below, the
bed should under no circumstances be put into service again.
Contact the supplier or manufacturer in this case.
Summary of Contents for WOBURN COMMUNITY 1200
Page 41: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 41 DECLARATION OF CONFORMITY ...
Page 43: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 43 NOTES ...
Page 44: ...T 44 0 1709 377172 F 44 0 1709 377173 44 NOTES ...
Page 45: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 45 NOTES ...
Page 46: ...T 44 0 1709 377172 F 44 0 1709 377173 46 NOTES ...
Page 47: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 47 ...