8. Error messages
Error which has
occurred
Possible causes
Remedy
Monitor cannot be
switched on
Batteries are missing, inserted incorrectly or flat.
Check batteries and insert two identical new batter-
ies if necessary.
The measuring signals were not read or were read
incorrectly. This can be caused by the cuff not
being applied correctly, moving, talking or a very
weak pulse.
Check that the cuff has been positioned correctly.
Do not talk or move during the measuring proce-
dure. Also observe the 10 golden rules.
E2
Invalid measurement due to movement.
Do not talk or move during the measuring proce-
dure.
E3
The cuff was not applied correctly.
Check that the cuff has been positioned correctly.
E4
Error during the measurement.
Contact customer service if this error message
appears.
E5
Inflation pressure is higher than 300 mmHg.
Please rest for at least 1 minute and take the meas-
urement again.
E6
There is a system error.
Contact customer service if this error message
appears.