![Hart Sceintific DewK 1620 User Manual Download Page 95](http://html1.mh-extra.com/html/hart-sceintific/dewk-1620/dewk-1620_user-manual_601476095.webp)
8.4.1.12
ALARm:POWer:ENABle?
Returns the state of the power loss alarm enable.
Example: ALAR:POW:ENAB?
Response: 0
A value of 1 is returned if the power loss alarm is enabled. A value of 0 is re-
turned if the power loss alarm is disabled.
8.4.1.13
ALARm:POWer:ENABle <bool>
Sets the state of the power loss alarm enable.
Example: ALAR:POW:ENAB 1
The <bool> parameter turns the power loss alarm on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.4.10, Pass-
word Commands).
8.4.1.14
ALARm:RHUMidity<chn>:LOWer?
Indicates whether a lower humidity alarm event has occurred for the specified
channel.
Example: ALAR:RHUM1:LOW?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if a lower humidity alarm event has occurred on the specified channel.
A value of 0 is returned if a lower humidity alarm event has not occurred on the
specified channel.
8.4.1.15
ALARm:RHUMidity<chn>:LOWer:ENABle?
Returns the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if the lower humidity alarm is enabled. A value of 0 is returned if the
lower humidity alarm is disabled.
8.4.1.16
ALARm:RHUMidity<chn>:LOWer:ENABle <bool>
Sets the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB 1
88
Hart Scientific
8 Digital Communications Interface