![Hart Sceintific DewK 1620 User Manual Download Page 109](http://html1.mh-extra.com/html/hart-sceintific/dewk-1620/dewk-1620_user-manual_601476109.webp)
or
Response: 1,1,25.629,C,29.29,%,2,0,C,0,%,2003,9,16,11,1,42
This command is equivalent to the FETC? command.
8.4.2.17
READ? [<chn>]
Returns a comma delimited list of the most recent measurements for the speci-
fied channel or for both channels.
Example 1: READ? 1
Response: 25.576,29.30
or
Response: 1,1,25.629,C,29.32,%,2003,9,16,11,2,38
Example 2: READ?
Response: 25.582,29.32,26.341,37.96
or
Response: 1,1,25.629,C,29.29,%,2,0,C,0,%,2003,9,16,11,1,42
This command is equivalent to the FETC? command.
8.4.3
Data Commands
The data commands are used to access daily statistics or recorded
measurements.
8.4.3.1
DATa:DSTatistics:ENABle?
Returns the state of the daily statistics recording enable.
Example: DAT:DST:ENAB?
Response: 0
A value of 1 is returned if the daily statistics recording is enabled. A value of 0
is returned if the daily statistics recording is disabled.
8.4.3.2
DATa:DSTatistics:ENABle <bool>
Sets the state of the daily statistics recording.
Example: DAT:DST:ENAB 1
The <bool> parameter turns the daily statistics recording on (1 or ON) or off (0
or OFF). This command may be password protected (see Section 8.4.10, Pass-
word Commands).
102
Hart Scientific
8 Digital Communications Interface