You will now see the above screen, instructing to either
<F1>Drain
or
<F4>Fill
the cuvette. If the cuvette has any residual fluid remaining, press and hold the
<F1>Drain
button until it has emptied completely. When cuvette is completely
empty, press the
<F4>Fill
button, this will set the “empty” level for the cuvette.
You will now be presented with the above screen. The fill level is set
individually for the Stain, and together for the Rinse/Buffer.
NOTE
: When setting the Stain and Rinse/Buffer levels, you must
press
and hold
the button down rather than a press-and-release.
Begin with the setting the Stain calibration. With the slide still in cuvette,
press
and hold
the
<F2>Stain
button until the cuvette fills up to the desired level.
Release
the
<F2>Stain
button when the desired Stain fill level has been met.
Once you release the button, the screen will then revert back to the
<F1>Drain
or
<F4>Fill
screen below.
The cuvette will currently be filled with stain.
Press and hold
the
<F1>Drain
to drain the cuvette until it is empty. You will see the stain drain from the waste
line (hypotube #3) underneath the cuvette. Once the cuvette is completely
empty, press the
<F4>Fill
button to proceed to set the rinse/buffer level.