9.0
Unit Installation
If you are installing your instrument for the first time, please use the following guide to
ensure proper set-up. For this setup, the unit will need to be powered off, and you will need to
have a HemaPRO Reagent Kit on hand.
Remove the two thumb screws to open the front panel that is covering the
pump tube rollers.
Attach all four orange pump tubes around their pump rollers.
Open the HemaPRO Reagent Kit (Cat. No.
HP1SK
). Remove the cap from
each reagent bottle, and pierce the induction seal on the opening of each bottle
with a sterile needle.