Unscrew the two (2) thumbscrews holding the front panel pump cover in place.
This will expose the four pump tube rollers, the orange pump tubes, and the
reagent/waste lines.
NOTE:
Please use the above diagram to assist you when replacing the
tubes from the pump rollers. Notice that all “elbow” connection labeled as
“
Shorter
” connect to the bottom of the cuvette.
Remove the orange pump tubes off of each roller so that they are hanging off
of the roller, but still held inside of the black metal bracket.
Cut the four (4) zip-ties that are holding the clear tubing together. One (1) may
be found between Pump #3 and Pump #4, and three (3) are located along the
back of the instrument.
Pump #1
Rinse
Pump #2
Buffer
Pump #3
Waste
Pump #4
Stain
Shorter
Longer
Shorter
Longer
Shorter
Longer
Shorter
Longer
Bracket