2.2 Operating procedure 

2.2.1 Independence use of two adjustable outputs. 

2.2.1.1 Set the (13) and (14) switch to spring out position ( 

   

position). 

2.2.1.2  When  the  adjustable  output  is  used  as  C.V.  output,  first  should  rotate 

clockwise the C.C. adjustment (6) and (22) to Max, then turn on power switch (7), 

adjust  C.V.  adjustment  (5)  and  (23)  till  slave  &  master  DC  output  voltage  reach 

required voltage value, at this time, the C.C. indicator (8) and (18) light off. 

2.2.1.3  Used  as  C.C.  output,  after  turning  on  power  switch  (7),  first  rotate 

clockwise the C.V. adjustment (5) and (23) to Max, while rotate counterclockwise 

the  C.C.  adjustment  (6)  and  (22)  to  Min,  then  connect  the  required  load,  again 

rotate  clockwise  adjustment  (6)  and  (22)  till  output  current  reach  the  required 

current value. At this time the C.V. State indicator (9) and (19) go out and the C.C. 

state indicator (8) and (18) light on. 

2.2.1.4  Used  as  the  C.V.  output,  in  general  the  C.C.  adjustment  (6)  and  (22) 

should be set to Max, but for this unit, the current-limiting protection point can also 

be set arbitrarily. Setting procedure is: Turn on power, rotate counterclockwise the 

C.C. adjustment (6) and (22) to Min, then make the positive and negative output 

terminal in short connection and rotate clockwise the C.C. Adjustment (6) and (22) 

till  output  current  equal  to  the  required  current-limiting  protection  point  so  the 