Specifications
Chapter 8 Specifications
8.1 Technical Specifications
All specifications herein mentioned apply to the DSO5000 series oscilloscopes. Before checking
an oscilloscope from HANTEK to see if it complies with these specifications, make sure it meets
the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5 .
℃
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.
Oscilloscope Specifications
Horizontal
Sample Rate Range
500MS/s--1GS/s
Waveform Interpolation
(sin x)/x
Record Length
Maximum 1M samples per single-channel; maximum 512K
samples per dual-channel (4K,16K,40K optional)
SEC/DIV Range
DSO5102C
DSO5062B DSO5102B
DSO5202M
4ns/div to 40s/div, in a 2, 4, 8
sequence
2ns/div to 40s/div, in a 2, 4, 8
sequence
Sample Rate and
Delay Time Accuracy
±50ppm over any ≥1ms time interval
Delta Time Measurement
Accuracy
(Full Bandwidth)
Single-shot, Normal mode
± (1 sample in100ppm × r 0.6ns)
>16 averages
± (1 sample in 100ppm × r 0.4ns)
Sample interval = s/div ÷ 200
Position Range
DSO5102C
DSO5062B DSO5102B
20ns/div to 80μs/div
(-8div × s/div) to 40ms
200μs/div to 40s/div
(-8div × s/div) to 400s
DSO5202B
DSO5000 Series Digital Storage Oscilloscope User Manual 68