Basic Operation
Press this button to see the dialog for buzzer and time setups. Turn the knob V0
to select buzzer or time (with a black frame). Press V0 and the black frame turn
to be red. Turn V0 again to set ON/OFF or time. Press V0 again to exit or save
the setups.
Self Calibration:
The self calibration routine can optimize the precision of the oscilloscope to fit
the ambient temperature. To maximize the precision, you should perform the self calibration once
the ambient temperature changes by 5 or more. Follow the instructions on the screen.
℃
Tip: Press any menu button on the front panel to remove the status display and enter a
corresponding menu.
5.5.5CURSOR
Push the CURSOR button to display the Cursor Menu.
Options
Settings
Comments
Type
Off
Voltage
Time
Select a measurement cursor and display it.
Voltage measures amplitude while Time measures time and
frequency.
Source
CH1
CH2
MATH
REFA
REFB
Select a waveform to take the cursor measurement.
Use the readouts to show the measurement.
Select Cursor S
E
S indicates Cursor 1. E indicates Cursor 2.
A selected cursor is highlighted, which can be moved freely. Both
cursors can be selected and moved at the same time. The box
behind the cursor displays the location of the cursor.
Delta
Display the
difference
(delta)
between the
cursors.
Display the measurement in the box under this option.
Moving Cursors:
Press the key near Select Cursor to select a cursor and turn V0 to move it.
Cursors can be moved only when the Cursor Menu is displayed.
DSO5000 Series Digital Storage Oscilloscope User Manual 42