Basic Operation

Falling Edge

Display the falling edge.

Undo Setup

Let the oscilloscope recall the previous setup.

5.6.2Help

Push the HELP button to display the Help menu which provides  topics covering  all  the menu
options and controls of the oscilloscope. For detailed information about the Help system, refer to

Section 2.2

.

5.6.3Default Setup

When you push the DEFAULT SETUP button, the oscilloscope will display the CH1 waveform
and remove all the others. When you are at the default setup, press F1 to 

Undo

 

Preset

. Then the

oscilloscope returns to the status before default setup. The table below gives the options, buttons
and controls that change settings at default setup.

Menu or System

Option, Button or Knob

Default Setting

Acquire

(Three mode options)

Normal

Averages

16

Run/Stop

Run

Cursor

Type

Off

Source

CH1

Horizontal (amplitude)

±3.2div

Vertical (time)

±4div

Display

Type

Vectors

Persist

Off

Format

YT

Horizontal

Window Mode

Single-window

Trigger Knob

Level

Position

0.00s

SEC/DIV

200μs

Math

Operation

Source

CH1-CH2

Position

0div 

FFT

Vertical Scale

20dB

FFT Operation 
Source

CH1

Window

Hanning

FFT Zoom

X1

Measure

Source

CH1

DSO5000 Series Digital Storage Oscilloscope User Manual                                            46

Summary of Contents for DSO5000 Series

Page 1: ...DSO5000 Series Digital Storage Oscilloscope User Manual...

Page 2: ...ord 5 3 2 FUNCTIONAL CHECK 5 3 2 1 Power on the oscilloscope 5 3 2 2 Connect the oscilloscope 5 3 2 3 Observe the waveform 6 3 3 PROBE EXAMINATION 6 3 3 1 Safety 6 3 3 2 Use of Probe Check Wizard 7 3...

Page 3: ...IONAL KNOBS AND BUTTONS 43 5 8 SIGNAL CONNECTORS 43 CHAPTER 6 APPLICATION EXAMPLES 45 6 1 EXAMPLE 1 TAKING SIMPLE MEASUREMENTS 45 6 2 EXAMPLE 2 TAKING CURSOR MEASUREMENTS 47 6 3 EXAMPLE 3 ANALYZING IN...

Page 4: ...ERVICES AND SUPPORT 69 CHAPTER 10 GENERAL CARE AND CLEANING 70 10 1 GENERAL CARE 70 10 2 CLEANING 70 APPENDIX A HARMFUL AND POISONOUS SUBSTANCES OR ELEMENTS 71 Appendix B Index 72 DSO5000 Series Digit...

Page 5: ...l rights to modify this document without prior notice Please contact Hantek for the latest version of this document before placing an order Hantek has made every effort to ensure the accuracy of this...

Page 6: ...ted to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Connect the probe in a right way The probe ground lead is at...

Page 7: ...essible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a possible hazard to this product or other property 1 4 Symbols...

Page 8: ...us to environments or human bodies To avoid them being released outside and to minimize the waste of natural resources we suggest you reasonably call back this device to ensure proper recovery and rec...

Page 9: ...s help you save a lot of time in operation the Autoset function lets you detect sine and square waves automatically the Probe Check Wizard guides you to adjust the probe compensation and set the Probe...

Page 10: ...s Turn the HELP SCROLL knob to move the highlight from one link to another Push the Show Topic option button to display the topic corresponding to the highlighted link Press the Back option button to...

Page 11: ...45 to 440 Hz 3 1 2Power Cord Use only power cords designed for this product Refer to 8 2 Accessories for specific standards 3 2 Functional Check Follow the steps below to perform a quick functional ch...

Page 12: ...t connect the probe tip and reference lead to the PROBE COMP connectors There is a mark on the panel Probe COMP 5V 1KHz 3 2 3Observe the waveform Press the AUTOSET button and you should see within a f...

Page 13: ...K button to use the Probe Check Wizard and configure the probe option attenuation factor properly following menu prompts 3 4 Manual Probe Compensation Upon the first connection of a probe and an input...

Page 14: ...f the Probe attenuation option matches the attenuation of the probe As an alternative method to Probe Check you can push a vertical menu button such as the CH 1 MENU button and select the Probe option...

Page 15: ...bient temperature changes by 5 or more For a more accurate calibration please power on the oscilloscope and wait for 20 minutes until it has adequately warmed up To compensate the signal path disconne...

Page 16: ...e the setup each time before being closed and automatically recall the setup once being turned on Note If you modify the setup please wait for more than 5 seconds before turning off the oscilloscope t...

Page 17: ...gh the trigger circuitry Trigger Mode You can select the Auto or Normal mode to define how the oscilloscope acquires data when it does not detect a trigger condition Auto Mode performs the acquisition...

Page 18: ...acquisition mode the oscilloscope acquires several waveforms averages them and displays the resulting waveform You can use this mode to reduce random noise Equivalent Acquisition This kind of acquisit...

Page 19: ...edge The oscilloscope shows the horizontal scale as time per division in the scale readout Since all active waveforms use the same time base the oscilloscope only displays one value for all the active...

Page 20: ...utomatically in this mode As this measurement uses the waveform record points it is more precise than the graticule and cursor measurements Automatic measurements show the measurement results by reado...

Page 21: ...will give a quick overview of all control buttons and knobs on the front panel as well as the displayed information on the screen and relative testing operations The figure below illustrates the front...

Page 22: ...orks in auto mode and is acquiring waveforms in the absence of triggers The oscilloscope is acquiring and displaying waveform data continuously in scan mode The oscilloscope has stopped acquiring wave...

Page 23: ...g edge Edge trigger on the falling edge Video trigger with line synchronization Video trigger with field synchronization Pulse Width trigger positive polarity Pulse Width trigger negative polarity 13...

Page 24: ...nusable 5 2 Horizontal Controls Use the horizontal controls to change the horizontal scale and position of waveforms The horizontal position readout shows the time represented by the center of the scr...

Page 25: ...Then it positions the window to this mark for further observation Holdoff None Select this menu and turn the multi functional knob to adjust the trigger holdoff time within the range of 100ns 10s Sel...

Page 26: ...d turn this knob to scale the waveform whose magnification is up to 1000 Notes 1 For more information of the trigger holdoff see Section Trigger Controls 2 In single window mode press F0 to hide or sh...

Page 27: ...position on the screen Two channels correspond to two knobs 2 Menu CH1 CH2 Display vertical menu options turn on or off the display of channel waveforms Options Settings Comments Coupling DC AC Groun...

Page 28: ...is unnecessary to be displayed can be used as a trigger source or for math operations 3 VOLTS DIV Knob Control the oscilloscope to magnify or attenuate the source signal of the channel waveform The ve...

Page 29: ...g the FFT mode Follow the steps below 1 Push the AUTOSET button to display a YT waveform 2 Turn the VERTICAL POSITION knob to vertically move the YT waveform to the center zero division so as to ensur...

Page 30: ...zontally to 250 points but you can use the FFT Zoom function to expand the FFT spectrum so that you can clearly view the frequency components at each of the 1024 data points in the FFT spectrum Note T...

Page 31: ...at the YT waveform repeats all the time When the number of cycles is integral 1 2 3 the YT waveform starts and ends at the same amplitude and there are no discontinuities in the signal shape If the nu...

Page 32: ...y for each type of window You shall determine which one to choose according to the object you want to measure and the source signal characteristics Window Measurement Characteristics Hanning Periodic...

Page 33: ...ndwidth of the source waveform to lower than the Nyquist frequency Identify and ignore the aliased frequencies Use zoom controls and cursors to magnify and measure the FFT spectrum 5 3 1 6 Magnifying...

Page 34: ...sure at any zoom factor Push the CURSOR button choose the Source option and then select Math Press the Type option button to select between Amplitude and Frequency Click the SELECT CURSOR option to ch...

Page 35: ...tical midpoint between the peaks of the trigger signal 3 Force Trigger Used to complete an acquisition regardless of an adequate trigger signal This button becomes useless if the acquisition is alread...

Page 36: ...ithin a certain amount of time based on the SEC DIV setting The oscilloscope goes into the scan mode at 80ms div or slower time base settings In the Normal mode the oscilloscope updates the display on...

Page 37: ...on negative going sync pulses If the video signal contains positive going sync pulses use the Inverted Polarity option Pulse Width Trigger You can use it to trigger on aberrant pulses Options Setting...

Page 38: ...ible and accurate than the Edge trigger Options Settings Comments Slope Source CH1 CH2 EXT EXT5 Select the input source as the trigger signal Slope Rising Falling Select the slope type of signal Mode...

Page 39: ...signals through the trigger circuitry Options Settings Comments Swap Trigger Mode Auto Normal Select the type of trigger Channel CH1 CH2 Push an option such as CH1 select the channel trigger type and...

Page 40: ...ext Page Vertical V1 V2 Adjust the vertical window by setting two trigger levels Select this option and press F3 to choose V1 or V2 When Select the trigger condition Time 20ns to 10 0sec Press F4 to s...

Page 41: ...oscilloscope detects one trigger and when it is ready to detect another During the holdoff time the oscilloscope will not trigger For a pulse train the holdoff time can be adjusted to let the oscillo...

Page 42: ...ff Display or remove the reference waveform on the screen Press Next Page to recall the following menu Options Settings Comments Setups Operation Source Flash memory USB disk Store the current setups...

Page 43: ...Calculate the time of the first cycle Mean Calculate the arithmetic mean voltage over the entire record Pk Pk Calculate the absolute difference between the greatest and the smallest peaks of the enti...

Page 44: ...aking Measurements For a single waveform or a waveform divided among multiple waveforms up to 8 automatic measurements can be displayed at a time The waveform channel must stay in an ON displayed stat...

Page 45: ...orithm to interpolate points between sampled points to produce a complete waveform record 4K by default Peak Detect Use this mode to detect glitches within 10ns and to limit the possibility of aliasin...

Page 46: ...dware versions serial number and some other information about the oscilloscope Update Program Insert a USB disk with upgrade program and the disk icon at the top left corner is highlighted Press the U...

Page 47: ...SOR button to display the Cursor Menu Options Settings Comments Type Off Voltage Time Select a measurement cursor and display it Voltage measures amplitude while Time measures time and frequency Sourc...

Page 48: ...time length to display each displayed sample point Format YT XY YT format shows the vertical voltage in relation to time horizontal scale XY format displays a dot between CH1 and CH2 each time when a...

Page 49: ...to a USB device just like the screen capturing function of a computer 5 6 1AUTOSET Autoset is one of the advantages digital oscilloscopes have When you push the AUTOSET button the oscilloscope will i...

Page 50: ...the signal is similar to a sine wave the oscilloscope displays the following options Sine Wave Options Details Multi cycle Sine Display multiple cycles that have appropriate vertical and horizontal sc...

Page 51: ...s F1 to Undo Preset Then the oscilloscope returns to the status before default setup The table below gives the options buttons and controls that change settings at default setup Menu or System Option...

Page 52: ...pe Slope Rising Mode Auto Coupling DC When Trigger Swap CH1 Type Edge Slope Rising Mode Auto Coupling DC Level 0 00v CH2 Type Edge Slope Rising Mode Auto Coupling DC Level 0 00v Trigger OT Source CH1...

Page 53: ...f overtime of the overtime trigger and slope trigger select menu options and so on Easy to operate F7 Push this button in single window mode to switch between dotted line display and cross display Pus...

Page 54: ...he external trigger signal 3 Probe Compensation Voltage probe compensation output and ground used to electrically match the probe to the oscilloscope input circuit The probe compensation ground and BN...

Page 55: ...e 6 Triggering on a pulse width 7 Triggering on a video signal Observing triggers on video fields and video lines 8 Using Slope Trigger to capture a particular slope signal 9 Using Overtime Trigger to...

Page 56: ...E button to see the Measure menu 2 Turn the knob V0 to select the first unspecified option marked by red arrow press V0 or F6 to enter the submenu 3 Select CH1 for the Source option Then repeatedly pu...

Page 57: ...select a cursor If S is selected turn V0 to move Cursor S on the screen if E is selected turn V0 to move Cursor E if both are selected turn V0 to move them at the same time 5 Put Cursor S on the first...

Page 58: ...to select a cursor If S is selected turn V0 to move Cursor S on the screen if E is selected turn V0 to move Cursor E if both are selected turn V0 to move them at the same time 5 Place Cursor S on the...

Page 59: ...tton and select Fine Turn the VERTICAL POSITION knob to accurately separate the waveform into 5 divisions 5 Turn the VERTICAL POSITION knob to center the waveform Position the waveform baseline to 2 5...

Page 60: ...you may follow the steps below to analyze this signal Observing Noisy Signal 1 Press the ACQUIRE button to see the Acquire menu 2 Push the Type option button and select Real Time 3 Push the Peak Dete...

Page 61: ...n button and select Real Time 3 Push the Average option button 4 Push the Averages option button and adjust the number of running averages to watch the change in the waveform display Note Averaging re...

Page 62: ...u 4 Push the Peak Detect option button 5 Push the TRIG MENU button and select Rising for the Slope option Then adjust the trigger level properly 6 Push the SINGLE SEQ button to start the acquisition U...

Page 63: ...d VERTICAL POSITION knobs to properly scale the waveform display 10 Use the Lissajous s oscillographic method to observe and calculate the phase differences following the formula below As sin A B or C...

Page 64: ...pe option push F2 to select CH1 for the Source option turn the TRIGGER LEVEL knob to set the trigger level at the bottom of the signal 6 Push F6 to enter the next page Select the When option button an...

Page 65: ...Push F1 to select Video for the Type option 3 Push the Source option button to select CH1 push the Polarity option button to select Normal push the Standard option button to select NTSC 4 Push the Syn...

Page 66: ...ope Signal In many occasions we are not only concerned about the edge of the signal but also want to know the rise and fall times of the signal To better observe this kind of signals we bring in the s...

Page 67: ...1 to select OT for the Type option push the Polarity option button to select Normal push the Mode option button to select Auto push the Coupling option button to select DC 3 Turn the Trigger Level kno...

Page 68: ...By using the math function of the oscilloscope you can get a better representation of the waveform on the screen To observe this signal follow the steps below 1 Set the Probe option attenuation to 10X...

Page 69: ...eps to set up the Math menu and observe the subtracted waveform as shown in the figure below Those in pink are added waveforms 6 11 Example 11 Measuring Data Propagation Delay When you doubt that ther...

Page 70: ...UTOSET button to trigger a stable waveform display 5 Adjust the horizontal and vertical controls to optimize the waveform display 6 Push the CURSOR button to view the Cursor menu 7 Push the Type optio...

Page 71: ...Application Examples DSO5000 Series Digital Storage Oscilloscope User Manual 66 CS DATA...

Page 72: ...may press the Auto Measure button to perform an automatic detection of signals at first Contact HANTEK Technical Support department in time if there is still no display of waveforms 3 If the waveform...

Page 73: ...nteed unless noted typical Oscilloscope Specifications Horizontal Sample Rate Range 500MS s 1GS s Waveform Interpolation sin x x Record Length Maximum 1M samples per single channel maximum 512K sample...

Page 74: ...mode 5V div to 10mV div 4 for Normal or Average acquisition mode 5mV div to 2mV div DC Measurement Accuracy Average Acquisition Mode Measurement Type Average of 16 waveforms with vertical position at...

Page 75: ...igger Level Range Source Range CH1 CH2 8 divisions from center of screen EXT 1 2V EXT 5 6V Trigger Level Accuracy typical Accuracy is for signals having rise and fall times 20ns Source Accuracy CH1 CH...

Page 76: ...the set slope Not Equal The oscilloscope triggers when the waveform slope is not equal to the set slope Less than The oscilloscope triggers when the waveform slope is less than the set slope Greater t...

Page 77: ...ng DC AC or GND Input Impedance DC coupled 1M 2 in parallel with 20pF 3pF Probe Attenuation 1X 10X Supported Probe Attenuation Factors 1X 10X 100X 1000X Maximum Input Voltage Overvoltage Category Maxi...

Page 78: ...9 8 40 to 71 Cooling Method Convection Humidity 104 or below 40 or below 90 relative humidity 106 to 122 41 to 50 60 relative humidity Altitude Operating and Nonoperating 3 000m 10 000 feet Random Vib...

Page 79: ...power cord special for this product In addition to the power cord shipped with your instrument you may purchase another one certified for the country of use A warranty card When there appears somethi...

Page 80: ...e 3 Contact HANTEK headquarters in China Headquarters HANTEK Technologies Co Ltd http www hantek net Address 5F NO 177 zhuzhou road huite industry city QingDao China 266000 Tel 86 532 88703687 8870369...

Page 81: ...oscope and probes as often as operating conditions require To clean the exterior surface perform the following steps 1 Use a lint free cloth to remove floating dust on the outside of the oscilloscope...

Page 82: ...0 Fastener and Installed Hardware X 0 X X 0 0 Other Accessories including probes X 0 0 X 0 0 Others 0 0 0 0 0 0 X means that at least the content of this poisonous and harmful substance in a homogene...

Page 83: ...se Fine 20 Connector 5 Contrast 39 Coupling 9 CURSOR 12 CURSOR button 12 Cyc RMS 34 D DISPLAY button 38 Dual window 17 E Edge trigger 10 EXT TRIG 28 F Factory setup 9 FFT aliasing 22 FFT Window 21 FFT...

Page 84: ...utton 36 S Sample rate 3 SAVE RECALL 33 Scale 11 Scan mode display roll mode 19 SECAM 28 SEC DIV knob 11 Self calibration 8 SET TO 50 button 61 Simple measurement 12 Sine wave 40 SINGLE SEQ button 10...

Page 85: ...Harmful and Poisonous Substances or Elements X XY format 16...

Reviews:

Related manuals for DSO5000 Series