background image

6

A100h 
JANUARY 2020

SENSOR ELEMENT

Typical sensor element life is based on normal operating 

conditions. Exposure to the target gas will shorten 

these times. The typical sensor element life of the 

electrochemical gas sensor element is 2 to 3 years; 

typical sensor life for semiconductor element is 4 to 6 

years; and typical sensor life for infrared element is 5 

to 7 years. Recalibration should be performed at least 

every 6 months. Calibration can be performed locally 

at detector installation site. If desired results cannot be 

achieved or if signal reads <1.2mA >2.1mA, steady 3mA, 

replacement of the sensor element may be required. 
 

ELECTROCHEMICAL SENSOR ELEMENT 

Electrochemical sensor element measure the partial 

pressure of gases under atmospheric conditions. The 

monitored ambient air diffuses through a membrane 

into the liquid electrolyte in the sensor. The electrolyte 

contains a measuring electrode, a counter-electrode and 

a reference electrode. An electronic “potentiostat” circuit 

ensures a constant electrical voltage between measuring 

electrode and reference electrode. Voltage, electrolyte, 

and electrode material are selected to suit the gas being 

monitored so that it is transformed electrochemically on 

the measuring electrode and a current flows through the 

sensor element. This current is proportional to the gas 

concentration. At the same time, oxygen from the ambient 

air reacts at the counter electrode electrochemically. 

The current flowing through the sensor is amplified 

electronically, digitized and corrected for several 

parameters (e.g., the ambient temperature).

SEMICONDUCTOR SENSOR ELEMENT 

Semiconductor or metallic oxide sensor element (MOSs) 

are among the most versatile of all broad-range sensor 

element. They can be used to detect a variety of gases 

and vapors in low ppm or even combustible ranges. 

The sensor element is made up of a mixture of metallic 

oxides. They are heated to a temperature between 150° 

and 300° C depending on the gas(es) to be detected. 

The temperature of operation as well as the “recipe” of 

mixed oxides determines the sensor element selectivity 

to various toxic gases, vapors, and refrigerants. Electrical 

conductivity greatly increases as soon as a diffusion 

process allows the gas or vapor molecules to come in 

contact with the sensor element surface. Water vapor, 

high ambient humidity, temperature fluctuations, and low 

oxygen levels can result in higher readings. 

INFRARED SENSOR ELEMENT

The infrared (IR) gas sensor element is designed to 

measure the concentration of combustible gases and 

vapors in the ambient air. The principle is based on 

the concentration-dependent absorption of infrared 

radiation in measured gases. 
The monitored ambient air diffuses through a sintered 

metal material into the enclosure of an optical “bench”. 

The broadband light emitted by an IR source passes 

through the gas in the optical bench and is reflected 

by the walls from where it is directed towards a dual-

element detector. One channel of the detector measures 

the gas-dependent light transmission, while the other 

channel is used as a reference. The ratio between 

measurement and reference signal is used to determine 

the gas concentration. Internal electronics and software 

calculate the concentration and produce an output signal

IMPORTANT:

 Certain substances in the atmosphere to 

be monitored may impair the sensitivity of the sensors. 

Such substances include, but are not limited to:
     • Polymerizing substances such as acrylonitrile,   

        butadiene and styrene.
      • Corrosive compounds such as halogenated  

         hydrocarbons (releasing halogens such as   

         bromine, chlorine or fluorine when oxidized)  

         and halogen hydride acids as well as acidic  

         gaseous compounds such as sulfur dioxide and  

         nitrogen oxides.
     • Catalyst poisons such as sulfurous and  

        phosphorous compounds, silicon compounds  

        (especially silicones), and metal-organic vapors.

LED LOGIC  

The Hansen gas detectors provide external indication 

of their current operational state via audible visual 

feedbackand also provide relays outputs. Visual 

indication of the instrument status is provided by 

a single tri-color LED (Green I Red I Orange) as 

indicated below: 

WIRING GAS DETECTORS 

Install the Hansen Gas Detector in an area where 

operating personnel can easily monitor it. Refer to 

Gas Detector Location

 on page 3 for suggestions on 

proper placement of Remote Detectors.
Use two-core shielded pair, 16 to 20 AWG wire to connect 

power to the gas detector. Connect the power input 

wires to terminal 4. Connect the analog signal wire to 

terminal 6 and/or digital sign (MODBUS) to terminal 5. The 

analog 4-20mA output signal is best for long distances 

to the monitor and where electrical noise is a problem. 

The voltage output is for short distances within 10 feet 

(3 m) of the monitor. The maximum 4-20mA cable length 

is 1000 ft (300 m).
Under no circumstances should the gas detector low 

voltage signal wires be in a common conduit, tray or 

wiring panel with power wiring over 48 volts. Do not run 

wires near variable frequency drive (VFD) equipment.  
Hansen recommends backup of gas detection system 

with an uninterruptable power supply to provide battery 

backup in the event of a power failure.

4.1.4 Status Indication 

The MGS-400 gas detectors provide external indication of their current operational state via 

audible and visual feedback. 

(MGS-450 I 460 gas detectors also provide relays outputs.) 

Visual 

indication of the instrument status is provided by a single tri-color LED 

(Green I Red I Orange)

as indicated below: 

Warm-up 

e

<I 

OFF 

OFF 

OFF 

Normal 

<I

OFF 

OFF 

OFF 

Low Alarm 

.))) 

<J,)) 

ON 

OFF 

OFF 

-

High Alarm 

e»») 

<J,� 

ON 

ON 

OFF 

Offline 

<I 

OFF 

OFF 

OFF 

Fault 

OFF 

OFF 

ON 

-

Negative Gas Fault 

I

e))))) 

OFF 

OFF 

I

ON 

Zero Cal. Fault 

I

.

))))) 

<I 

OFF 

OFF 

OFF 

I Span Cal. Fault 

»))) 

<I 

OFF 

OFF 

OFF 

4.1.5 Switch Functions 

User interaction with the MGS-400 gas detector is accomplished through the use of two 

magnetic switches located on the bottom of each unit. To actuate a magnetic switch 

(referred to

as MAG#1 or MAG#2), 

apply the supplied magnetic wand 

(PIN: 1100-1004) 

to the relevant 

switch location as indicated below: 

� 

C, 

Cl 

BACHARACH 

1100-2294 Rev 1 

24 

TABLE  3

Summary of Contents for HGD-EC-NH3-100

Page 1: ...sent in the surrounding environment The analog outputs 4 20mA or 0 10V DC can interface with nearly any existing monitor computer or PLC controller The sensor elements are mounted internally on the en...

Page 2: ...120 F 50 C 4 F 20 C to 122 F 50 C 13 F 25 C to 131 F 55 C VOLTAGE REQUIREMENTS 24 VDC 18 24 VAC 20 100 240 VAC 50 60 Hz 80W max D Cell BaVery HUMIDITY RANGE NON CONDENSING 5 90 RH 15 90 RH electroche...

Page 3: ...e most likely sources of leakage Perimeter detection is where detectors completely surround the area in question The size and nature of the area will help to decide which method is the most appropriat...

Page 4: ...D AVOID FALSE ALARMS REDUCING BUSHING SCREEN STAND ALONE CONTROL The Hansen Gas Detector can operate as a stand alone device All that is needed is a 24V AC DC power supply LOCAL AND REMOTE VISUAL AUDI...

Page 5: ...nly 3 Internal Alarm Buzze 4 Power Connections x2 5 Digital Connection MODBUS 6 Analog Connection 7 Tactile Switch 1 8 Ribbon Cable to Sensor 9 Tactile Switch 2 10 Relay 3 Connection FAULT 11 Relay 2...

Page 6: ...f the detector measures the gas dependent light transmission while the other channel is used as a reference The ratio between measurement and reference signal is used to determine the gas concentratio...

Page 7: ...el or branching multiple units from the main bus may introduce impedance mismatches reflections and or signal distortions Avoid long stubs when connecting detectors to the bus should be less than 3 De...

Page 8: ...is proportional to the detected gas concentration GAS CONCENTRATION 4 20mA 1 SV 0 SV 2 l0V 0 l0V 0 4mA lV ov 2V ov 50 12 mA 3V 2 5V 6V 5V 100 20 mA 5V 5V l0V lOV TABLE 6 FIG 11 8b Zero Adjustment 9 Bu...

Page 9: ...n user would need to download the Hansen Gas Detector App from either Google Play Store or Apple Store SENSOR SETUP In order to utilize the Bluetooth option sensors will need to be setup individually...

Page 10: ...verify the instrument functionality The steps to perform a bump test with the Hansen Gas Detector APP 1 Connect adapter and gas cylinder according to the instructions in the General Calibration Proce...

Page 11: ...ogrammable Logic Controller PLC Data logging is available on the HGD C 8 via the integrated SD card which can be removed to allow download of the logged data to a computer External Component Descrip2o...

Page 12: ...scrip on 1 Integrated Visual Alarm 2 Controller Power and Fault LEDs 3 Channel 1 and 2 alarm mute bu ons 4 Power Fault Low High Alarm LEDs 5 M16 Cable Glands x6 6 M20 Cable Glands x2 Component Descrip...

Page 13: ...UARY 2020 INSTALL DIMENSIONS 8 CHANNEL CONTROLLER FIG 18 5 625 0 0 0 6 0 O 0 ____ __ 16 J 1 47 REFERENCE NOT TO SCALE REFERANCE NOT TO SCALE 47 0 13 2 8 0 16 5 625 6 INSTALL DIMENSIONS 2 CHANNEL CONTR...

Page 14: ...14 A100h JANUARY 2020 FIG 20 CONTROLLER WIRING CONNECTIONS...

Page 15: ...he currently selected menu item is indicated with triangle pointers to the left and right of the description If the menu list is longer than can be displayed up and down arrow keys on the right side o...

Page 16: ...nec ng you must ENABLE Bluetooth from this menu screen From the LCD contrast seqng screen use the arrow keys to adjust the contrast from 1 63 with 30 being the default The brightness of front panel LE...

Page 17: ...MS device non Amer pted to e to edit ord a 30 s amer enter n 0 seqngs ma on nt since re rd bring assigned accessed either by selec ng CHANNEL CONFIG from the main menu or by pressing the channel numbe...

Page 18: ...ttings of the Hansen gas detector must match the BMS system to work correctly The 3 form C relays that are included in the Hansen gas detector fault low alarm high alarm can be configured to be fail s...

Page 19: ...D display open the clear cover and remove the LED display panel by removing the 4 screws On the backside of the LED display adjust the poten ometer labeled Z un l the KEY FEATURES GAS ALERT SYSTEM Pro...

Page 20: ...N100 etc WIRING SCHEMATIC FOR CONTROLLER AND DETECTOR External Power Supply EARTH GRD CUSTOMER BMS 0 J w w I i i z i 0 0 n r7 r7 r7 I i t i t ppm L J L J L J I 0 w I i i z 0 0 n 0 0 i 0 0 0 z 5 5 w w...

Page 21: ...connected to relay _______________________________ OK Check Analog Output in Use e g 0 5V 4 20 mA _______________________________ OK 2 On Site Gas Calibration Initial the following encircle OK when co...

Page 22: ...EP1 Handheld unit without sensor module HGD C 2 31 1012 Electrochemical 0 50 500 PPM HGC C 8 8 Channel controller 115 230 VAC 31 1013 Electrochemical 0 500 2000 PPM 31 1032 Filter 10 pack CALIBRATION...

Reviews: