40
Edit Standard Group
If a
Group Standard
was selected in the parameter Manual Entry,
this parameter is used to create your group of standards. If the
Standard Group already contains five ISE standards, at least one
ISE standard has to be removed in order to add another standard.
To edit/set the Standard Group:
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the Calibration option.
• Press
and use
or
to highlight the Edit
Standard Group option.
• Press
and use
and
to choose the ISE standard to be included in the standard group.
• Press
/
to add/remove the selected ISE standard to/from the standard group.
• Press
to return to Calibration options and to save the changes.
Calibration Reminder
- See Calibration option from pH Setup section.
Set Reminder Period
- See Calibration option from pH Setup section.
Clear Calibration
- See Calibration option from pH Setup section.
Electrode Type
This option allows the user to select the desired Ion Selective Electrode used for measurements from a list:
Ammonia, Bromide, Cadmium, Calcium, Carbon Dioxide, Chloride, Cupric, Cyanide, Fluoride, Iodide, Lead,
Nitrate, Potassium, Silver, Sodium, Sulfate, Sulfide and five custom ISE. For the standard ISE it is possible to view
the Ion constants (Name, Molar Weight and Electric Charge/Slope), while for the custom ISE all these constants
can be manually set.
41
To set the Electrode Type:
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the Electrode Type option.
• Press
and use
or
to select the desired
standard ISE or a custom one from the list.
For standard ISE:
• Press
to visualize the Ion constants and then press
at any time to exit Ion Constants view mode.
• Press
to confirm your selection and return to
ISE Setup
options.
For custom ISE:
• Press
to edit the Ion constants for the selected custom
ISE. Use
or
to select the desired constant and
press
to enter edit mode or
to cancel operation.
• For the Ion name the Text Editor menu will be displayed on the
LCD. Enter the desired information by accepting the highlighted
character which is added in the text bar, using
. The
and
keys help the user to select the desired
character. It is also possible to delete the last character by
positioning the cursor on the Backspace character and pressing
. Press
to return to the Ion Constants menu. If
the Saving Confirmation is enabled, press
to accept the
modified option,
to escape without saving or
to return to the editing mode. Otherwise, the modified option is
saved automatically.
• To set the appropriate Ion molar weight (in g/mol units) use
or
to set the desired value
and then press
to save the current value or press
to cancel
operation.