mV SETUP

The mV Setup menu allows the user to set the parameters associated with mV and Relative mV
measurements. These parameters can be set specifically for each channel (

HI 5222

 only). The settings will

be applied only to the active channel.

Accessing mV Setup

• Press 

 while in 

Measure

 mode and then 

 or

 to select mV / Rel mV range for the desired

channel.

• Press 

 and then 

 to access mV Setup menu.

To access a mV Setup option:

• Use 

 or 

 to highlight the desired option.

• Press 

 to access the selected option.

The following is a detailed description of the mV Setup option
screens.

Profile

See pH Setup section.

Temperature

ORP measurements are not temperature compensated, although
ORP values can change with temperature (e.g. reference electrode
potential changes, sample equilibrium changes). It is important to
report ORP values together with the reference electrode used and the
temperature of measurement.
This option permits selection of the temperature source (

HI 5222

only) and measurement units.

47

Verify the pH electrode and instrument has been calibrated before making pH measurements.

DIRECT MEASUREMENT

To measure the pH of a sample using the Direct reading mode:

• Press 

 and then 

 to select 

pH Measure

mode (select channel first for 

HI 5222

).

• Select the Direct reading mode (see pH Setup for details).
• Place the electrode tip and the temperature probe

approximately 4 cm (1.5”) into the sample to be tested.
Allow time for the electrode to stabilize. “

Stable

” will

appear.

• The measured pH value will be displayed on the LCD,

together with a short GLP information and display preferences.

Note

: If the reading is out of range, “

----

” will be displayed

on the LCD.

DIRECT/AUTOHOLD MEASUREMENT

To measure pH of a sample using the Direct/AutoHold reading
mode:

• Press 

 and then 

 to select 

pH Measure

mode (for the selected channel - 

HI 5222

 only).

• Select the Direct/AutoHold reading mode (see pH Setup for

details).

• Place the electrode tip and the temperature probe

approximately 4 cm (1.5”) into the sample to be tested.

• The measured pH value will be displayed on the LCD. Press

 and the “

AutoHold

” indicator will start blinking

on the LCD until the stability criterion is reached. The pH
value will be frozen on the LCD, along with “

AutoHold

” indicator.

Unrecognized buffer. Please check the buffer or the buffer list 

(for Semiautomatic and Automatic

buffer entry type): this message appears if the current buffer value is not close to any of the buffers from
the buffer list/group. Check if the current buffer is present in the buffer list or the appropriate buffer group
was selected.

The current buffer was already calibrated:

 change the buffer or press 

 to exit calibration

mode.

pH MEASUREMENT

Summary of Contents for HI 5221

Page 1: ...i n s t c o m MAN5222 02 15 PrintedinROMANIA Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 757...

Page 2: ...e mail us at tech hannainst com or view our worldwide contact list at www hannainst com WARRANTY All rights are reserved Reproduction in whole or in part is prohibited without the written consent of t...

Page 3: ...king material and examine it carefully to make sure that no damage has occurred during shipping If there is any damage notify your Dealer or the nearest Hanna Customer Service Center PRELIMINARY EXAMI...

Page 4: ...HI 5222 same for each measurement channel Log files include GLP information Basic Measurement can be viewed with detailed GLP information or with a Graph or a Log History User friendly interface on la...

Page 5: ...age Profile feature store up to ten different user setup HI 5222 five on each channel 76 HI 2031B Glass body semimicro conical refillable combination pH electrode Use semisolid products HI 1332B Plast...

Page 6: ...nnel 2 10 Temperature probe socket Channel 2 11 BNC electrode connector for pH ORP ISE measurements Channel 2 REAR PANEL 75 HI 1131B Glass body refillable double junction combination pH electrode Use...

Page 7: ...eral information about the selected option operation 74 ORP SOLUTIONS HI 7020L Test Solution 200 275 mV 500 mL bottle HI 7021L Test Solution 240 mV 500 mL bottle HI 7022L Test Solution 470 mV 500 mL b...

Page 8: ...p m I t u p n I 0 1 2 1 s m h o y a l p s i D s l e x i p 0 2 3 x 0 4 2 D C L c i h p a r G r o l o C y l p p u S r e w o P r e t p a d a c d V 2 1 t n e m o r i v n E g n i s n e d n o c n o n H R 5...

Page 9: ...utton located on the rear panel of the instrument Please wait until the instrument finishes the initialization process Note It is normal for the loading process to take a few seconds If the instrument...

Page 10: ...n i e r a n o i t c n u j e c n e r e f e r e h t n i s i e l p m a s e h t e r u s e k a M e h t k c e h C e g n a r d e i f i c e p s l a r e n e g e h t d n a l e v e l e t y l o r t c e l e E S I...

Page 11: ...r a faster response and to avoid cross contamination of the samples rinse the electrode tip with a few drops of the solution to be tested before taking measurements STORAGE PROCEDURE To minimize clogg...

Page 12: ...he protective cap off the pH electrode SALT DEPOSITS MAY BE PRESENT This is normal with electrodes They will disappear when rinsed with water During transport tiny bubbles of air may form inside the g...

Page 13: ...emperature During calibration the instrument will display the pH buffer value at 25 C P M E T S R E F F U B H p C K F 9 7 6 1 0 0 0 3 0 1 0 4 2 6 8 6 0 1 0 7 7 7 1 9 0 1 0 0 1 4 5 4 2 1 0 3 7 2 2 3 0...

Page 14: ...MP the sample concentration CSTD the standard concentration VSAMP the sample volume VSTD the standard volume VISA ISA volume E the difference of potential from the electrode S the electrode slope dete...

Page 15: ...adjustment is made to samples and standards Unknowns are measured directly by the instrument At lower concentrations in non linear regions of the electrode response multiple calibration points will e...

Page 16: ...e Press 65 ADDITIONAL INFORMATION ISE THEORY An Ion Selective Electrode ISE is an electrochemical sensor that changes voltage with the activity or concentration of ions in solutions The change in volt...

Page 17: ...ion Press and use key to highlight the desired parameter Use or to adjust the contrast backlight or to set the desired backlight saver time Press to confirm the modified options and return to the Syst...

Page 18: ...ss and then the desired parameter or HI 5222 only Only the selected measurement parameter lots will be displayed on the LCD Select the desired lot with or and press to display the logged data from the...

Page 19: ...meter reaches the stability criteria to go into Auto Hold mode The stored sample logs will be marked with an H to indicate the Auto Hold mode The virtual key returns operation to real time continuous...

Page 20: ...e r i D 2 l a u n a M g o L l a u n a M A N d l o H o t u A e l b a c i l p p A t o N 3 c i t a m o t u A g o L c i t a m o t u A d l o H o t u A t c e r i D 4 l a u n a M g o L l a u n a M 5 d l o H...

Page 21: ...p of the meter and ensures that the same procedure will be used To save the measurement configuration for pH mode Press and use or to highlight Profile Press to enable disable this feature pH SETUP 60...

Page 22: ...elect the Analyte Subtraction method see ISE Setup Reading Mode Prior to starting an AS procedure the ISE sensor must be calibrated with a minimum of two standards containing ISA The slope of the elec...

Page 23: ...he best results As a general rule the standard must be less concentrated than the sample so the addition of sample will increase the total Ion content of the beaker and change the mV value by at least...

Page 24: ...the observed mV Using a micropipet add a volume of the Reactant standard Start by adding 500 L at a time for example Watch the change in mV When you have observed approximately a 15 mV change from the...

Page 25: ...the solution and a mV value will show on the display Press to take the first mV reading When the reading is stable press to store the first mV reading The second step of the method will be displayed o...

Page 26: ...od Press and then to select ISE Measure mode for the selected channel Select the Known Addition method see ISE Setup for details Prior to starting a KA procedure the ISE sensor must be calibrated with...

Page 27: ...e of the incremental methods for measurement at least a two point ISE calibration must be performed to establish the electrode slope For accurate measurements add the appropriate ISA Ionic Strength Ad...

Page 28: ...oosing the AutoHold option the current reading can be frozen on the LCD when is pressed and the stability criterion is reached 53 If the isopotential point of the electrode is unknown the ISE calibrat...

Page 29: ...ared by pressing After 10 seconds will no longer be available Note It is very important to clear calibration history when a new electrode is used because most errors and warning messages that appear d...

Page 30: ...BRATION HI 5222 only For greater accuracy it is recommended to calibrate the ISE sensors frequently The instrument should be recalibrated when ISE x Calibration Expired the x represents channel 1 or c...

Page 31: ...larm State Press while in pH Measure mode Press Use or to select the Alarm option Press anduse or tohighlighttheAlarm State option Press anduse or tohighlightthedesired option Press to confirm your se...

Page 32: ...UREMENTS mV ORP MEASUREMENTS Oxidation reduction potential ORP measurements provide the quantification of the oxidizing or reducing power of the tested sample To correctly perform a redox measurement...

Page 33: ...le solution The pH reading is affected by temperature In order to measure the pH accurately the temperature effect must be compensated To use the Automatic Temperature Compensation ATC feature connect...

Page 34: ...for HI 5222 Select the Direct reading mode see pH Setup for details Place the electrode tip and the temperature probe approximately 4 cm 1 5 into the sample to be tested Allow time for the electrode...

Page 35: ...value can be adjusted using or Press to save the new temperature value If using Custom buffers press after buffer has been accepted to change actual buffer conditions A pop up menu will be displayed o...

Page 36: ...ctive nature of stability Known Addition In the Known Addition method a sample is measured with an ISE before and after the addition of a known volume of a standard The mV difference is then used to c...

Page 37: ...trode Cond Unknown pH x Default Calibration or pH x Calibration Expired message appears on the LCD in the Reminder messages area The x represents channel 1 or channel 2 PREPARATION Pour small quantiti...

Page 38: ...benefit from temperature compensated corrections if standards and sample temperatures differ from each other the Isopotential Point of the ISE is known If sample and standards are made at the same tem...

Page 39: ...key to select the next custom standard to be set Press to return to Edit Custom Standard options 42 To select the appropriate Ion Charge Slope use or and then press If the Ion electric charge is None...

Page 40: ...n constants Name Molar Weight and Electric Charge Slope while for the custom ISE all these constants can be manually set 41 To set the Electrode Type Press while in ISE Measure mode Press Use or to se...

Reviews: