21
METER SETUP
View Calibration Points
Option: Enable or Disable
With item selected, press the functional key to select desired option.
If enabled, the buffers used in the last calibration are displayed on the pH measurement screen.
Out of Calibration Range Warning
Option: Enable or Disable
With item selected, press the functional key to select desired option.
If enabled, when the pH reading is outside the calibration range, the “Out Cal. Range” message is
displayed.
ISE Probe
Option: 17 standard ISE probes or custom
Press
Standard
functional key to select desired probe from the standard probes list.
Press
Custom
functional key to set the parameters for a custom probe.
If
Standard
is pressed, press the / keys to select desired probe.
Press
Accept
to save or
ESC
key to return to the previous screen.
Press
View
to see probe parameters then
ESC
to return to the probes list screen.