METER SETUP
8.2. PROBE SPECIFIC PARAMETERS
Calibration Timeout
Option: Disabled, 1 to 7 days
With item selected, press
Modify
functional key. Press the / keys to set desired option. Press
Accept
to save or
ESC
key to return to the Meter Setup menu.
Note:
If enabled, “CAL DUE” warning will be displayed after the selected time has elapsed.
First Point Mode
Option: Replace or Offset
With item selected, press the functional key to select desired option.
With “Replace”, the slopes between current buffer and nearest lower and higher buffers will be
reevaluated.
With “Offset”, an electrode offset correction is performed keeping unchanged the existing slopes.
Custom Buffers
Option: None, 1 to 5
With item selected, press
Modify
functional key.
Press
Add
to add a custom buffer to the list (maximum 5).
With a buffer highlighted, press
Modify
to edit custom buffer value. Press the / keys to change
the value. Press
Accept
to save or
ESC
key to return to the custom buffer list screen.
Press
Delete
to delete the selected custom buffer from the list.
Press
ESC
key to return to the Meter Setup menu.