16
METER SETUP
The probe specific SETUP menu items are detailed here:
Item
Description
Option
Default
Calibration Timeout
(pH & ISE)
Number of days after
Calibration warning is
displayed
Disable, 1 to 7 days
Disable
First point mode (pH)
Management of
1 point calibration
Replace or offset
Replace
Custom buffer (pH)
Custom buffer setting
Max. 5 buffers
No
View calibration points
(pH)
Display calibration
points
Enable or disabled
Enable
Out of Cal. Range Warning –
Enable or disabled
Enable
ISE probe
Type of ISE probe
Custom or Standard (17)
Fluoride
ISE unit
–
User, ppt, g/L, ppm, mg/L,
ppb, µg/L, mg/mL, M,
mol/L, mmol/L, % W/V
ppm
8.1. GENERAL PARAMETER SCREENS
Backlight
Option: 0 to 7
This function allows the adjustment of the LCD backlight intensity.
With item selected press
Modify
functional key and use the arrow keys to change the backlight
intensity.
Press
Accept
to save or press
ESC
key return to the Meter Setup menu.