ELECTRODE CONDITIONING & MAINTENANCE

PREPARATION

Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT: this is
normal with electrodes and they will disappear when rinsed with
water.
During transport tiny bubbles of air may have formed inside the glass
bulb. The electrode cannot function properly under these conditions.
These bubbles can be removed by "shaking down" the electrode as
you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in 

HI 70300

Storage Solution for at least one hour.
For refillable electrodes, if the refill solution (electrolyte) is more than
2½ cm (1") below the fill hole, add the appropriate Electrolyte
Solution.

MEASUREMENT

Rinse the electrode tip with distilled water, immerse it (4 cm / 1½")
in the sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with the solution to be tested, before taking
any measurements.

STORAGE

To minimize clogging and ensure a quick response time, the glass
bulb and the junction should always be kept moist.
When not in use, store it with a few drops of 

HI 70300 

storage

solution or 

HI 7007

 pH 7.01 buffer in the protective cap.

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.

PERIODIC MAINTENANCE

Inspect electrode and cable. The cable used for the connection to the
meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb. If any
scratches or cracks are present, replace the electrode. Rinse off any salt
deposits with water.
Connectors must be perfectly clean and dry.
Refill the electrode with fresh electrolyte (see the electrode’s specifica-
tions to select the correct refilling solution). Allow the electrode to
stand upright for 1 hour. Follow the Storage Procedure.

• Wait till the hourglass symbol on the LCD turns off and the "pH"

tag stops blinking, then press CFM to confirm the second calibra-
tion point (slope).

• The meter automatically returns to normal measurement mode.

Note:

 To abort a calibration procedure and save previous calibration

data, press the CAL key while the "pH" tag is still blinking.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable inter-
ferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges.
Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the
electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the
instrument's EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measure-
ment surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.

Summary of Contents for pH 21

Page 1: ...pH20 pH21 Basic pH meters Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...struction manual carefully before using the instru ment This manual will provide you with all the necessary information for the correct use of the instrument as well as a precise idea of its versatili...

Page 3: ...EAR PANEL 1 BNC connector for pH or ORP pH 21 only electrode 2 Temperature probe socket for use with the HI 7662 3 Power supply input Plug the 12 Vdc adapter into the DC socket then into the mains Not...

Page 4: ...nking The LCD will show the ORP mV value TAKING pH MEASUREMENTS Make sure that the instrument has been calibrated for pH before taking measurements see pH Calibration section for details Connect the s...

Page 5: ...H 10 01 buffer solution The meter automatically recognizes the buffer and displays the correct value After measurements turn the meter off by pressing the ON OFF key disconnect the power adapter from...

Page 6: ...roken insulation on the cable or cracks on the electrode stem or bulb If any scratches or cracks are present replace the electrode Rinse off any salt deposits with water Connectors must be perfectly c...

Page 7: ...ectrodes HI 8071 3 5M KCl AgCl electrolyte 4 x 30 mL for single junction electrodes FDA approved bottle HI 7072 1M KNO3 electrolyte 4 x 30 mL CLEANING PROCEDURE General Soak in HI7061 General Cleaning...

Page 8: ...s HI 8082 3 5M KCl electrolyte 4 x 30 mL for double junction electrodes FDA approved bottle ORP SOLUTIONS HI 7091M Reducing pretreatment solution 230 mL bottle HI 7091L Reducing pretreatment solution...

Page 9: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Reviews: