7

6

Note

: If measurements are taken in different samples successively, it is

recommended to clean the electrode thoroughly to avoid cross-
contamination. After cleaning, rinse the electrode with water
and then with some of the solution to be tested.

MANUAL TEMPERATURE COMPENSATION

• When the temperature probe is not connected, the default

temperature value is 25°C.

• To temperature compensate the pH read-

ing, measure the temperature of the solution
with a 

ChecktempC

 or another accurate

thermometer. E.g. 20°C.

• Press the RANGE key to enter the °C mode and adjust the

displayed value by using the arrow keys.

TAKING ORP MEASUREMENTS (for pH21 only)

The meter is factory calibrated for ORP (mV) range. Please
contact the nearest Hanna Service Center for recalibration,
if needed.

• Connect an ORP electrode (optional, see "Accessories" section) to the

BNC connector on the rear panel (see page 4).

• Connect the 12VDC power adapter to the instrument (see page 4)

and then to the mains.

• Press the ON/OFF key to turn the meter

on.

• Press the RANGE key to select the ORP

(mV) mode.

• Remove the electrode protective cap, then immerse the ORP

electrode in the sample to be tested.

Note

: The electrode should be submerged approximately 4 cm (1½")

into the solution.

Note

: With refillable electrodes, to obtain a faster response unscrew

the refill hole cap during measurements.

• Shake briefly and wait a few seconds for

the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the ORP (mV) value.

TAKING pH MEASUREMENTS

• Make sure that the instrument has been calibrated for pH before

taking measurements (see "pH Calibration" section for details).

• Connect the supplied 

HI 1110B

 pH electrode to the BNC connec-

tor on the rear panel (see page 4).

• If automatic temperature compensation is required, connect the

HI 7662

 temperature probe (optional) to the proper socket on

the rear panel (see page 4).

• Connect the 12VDC power adapter to the instrument (see page 4)

and then to the mains.

• Press the ON/OFF key to turn the meter

on.

• Press the RANGE key to select the pH

mode, if necessary.

• Remove the electrode protective cap, then immerse the pH

electrode and the temperature probe in the sample to be tested.

Note

: If the pH electrode is dry, immerse it in 

HI 70300

 storage

solution for at least one hour to reactivate it.

Note

: The pH electrode should be submerged

approximately 4 cm (1½") into the
solution and the temperature probe (if
used) should be located as close as
possible to the electrode.

Note

: With refillable electrodes, to obtain a faster response unscrew

the refill hole cap during measurements.

• Shake briefly and wait a few seconds for

the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the pH value automatically com-
pensated for temperature variations.

• After measurements, turn the meter off by pressing the ON/OFF

key, disconnect the power adapter from the mains and the pH
electrode from the instrument.

• Store the electrode with a few drops of 

HI 70300 

solution in the

protective cap.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STORING
PURPOSES.

OPERATIONAL GUIDE

Summary of Contents for pH 21

Page 1: ...pH20 pH21 Basic pH meters Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...struction manual carefully before using the instru ment This manual will provide you with all the necessary information for the correct use of the instrument as well as a precise idea of its versatili...

Page 3: ...EAR PANEL 1 BNC connector for pH or ORP pH 21 only electrode 2 Temperature probe socket for use with the HI 7662 3 Power supply input Plug the 12 Vdc adapter into the DC socket then into the mains Not...

Page 4: ...nking The LCD will show the ORP mV value TAKING pH MEASUREMENTS Make sure that the instrument has been calibrated for pH before taking measurements see pH Calibration section for details Connect the s...

Page 5: ...H 10 01 buffer solution The meter automatically recognizes the buffer and displays the correct value After measurements turn the meter off by pressing the ON OFF key disconnect the power adapter from...

Page 6: ...roken insulation on the cable or cracks on the electrode stem or bulb If any scratches or cracks are present replace the electrode Rinse off any salt deposits with water Connectors must be perfectly c...

Page 7: ...ectrodes HI 8071 3 5M KCl AgCl electrolyte 4 x 30 mL for single junction electrodes FDA approved bottle HI 7072 1M KNO3 electrolyte 4 x 30 mL CLEANING PROCEDURE General Soak in HI7061 General Cleaning...

Page 8: ...s HI 8082 3 5M KCl electrolyte 4 x 30 mL for double junction electrodes FDA approved bottle ORP SOLUTIONS HI 7091M Reducing pretreatment solution 230 mL bottle HI 7091L Reducing pretreatment solution...