pH CALIBRATION

For better accuracy the instruments should be calibrated frequently,
and however:

1. After electrode replacement.
2. After testing aggressive chemicals.
3. At least once a month.
4. When extreme accuracy is required.

• Turn the instrument on after connecting the pH electrode and the

temperature probe.

• Immerse the temperature probe together with the electrode into

pH 7.01 buffer solution or manually set the temperature (see the
note here below).

Note:

 The pH electrode should be submerged approximately 4 cm

(1½") into the solution and the temperature probe should be
located as close as possible to the electrode.

Note:

 If the temperature probe is not used, measure the buffer value

with a thermometer, enter the °C mode by pressing the RANGE
key and adjust the displayed value by using the arrow keys.

• Press the RANGE key to display pH readings.

Press the CAL key to enter the calibration mode: the display
flashes "pH 7.01" and the hourglass symbol lights up.

• Wait till the hourglass symbol on the LCD turns off and the "pH"

tag stops blinking, then press CFM to confirm the first calibration
point (offset). "pH 4.01" will flash on the LCD.

• If a single point calibration at pH 7.01 is desired, press the CAL

key to exit, otherwise proceed with the second point.

• Rinse and immerse both pH electrode and temperature probe (if

used) in pH 4.01 or pH 10.01 buffer solution. The meter
automatically recognizes the buffer and displays the correct value.

• After measurements, turn the meter off by pressing the ON/OFF

key, disconnect the power adapter from the mains and the
electrode from the instrument.

Note

: To perform correct ORP measurements, the surface of the ORP

electrode must be clean and smooth.

Note

: When not in use, the tip of the electrode should be kept moist

and safe from any mechanical stress which might cause
damage to the glass/platinum junction.

TAKING TEMPERATURE MEASUREMENTS

The meter is factory calibrated for temperature range.
Please contact the nearest Hanna Service Center for
recalibration, if needed.

• Connect the 

HI 7662

 temperature probe (optional) to the proper

connector on the rear panel (see page 4).

• Connect the 12VDC power adapter to the instrument (see page 4)

and then to the mains.

• Press the ON/OFF key to turn the meter

on.

• Press the RANGE key to select the tem-

perature (°C) mode.

• Immerse the probe in the sample to be tested.
• Shake briefly and wait a few seconds for

the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the temperature (°C) value.

• After measurements, turn the meter off by pressing the ON/OFF

key, disconnect the power adapter from the mains and the probe
from the instrument.

HOLD FUNCTION

• To freeze a reading on display, from measurement mode press

and hold the HOLD key. The "HOLD" tag will light up.

• The meter returns to normal mode when the HOLD key is released.

Note

: The RANGE key is not working while in HOLD mode.

Summary of Contents for pH 21

Page 1: ...pH20 pH21 Basic pH meters Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...struction manual carefully before using the instru ment This manual will provide you with all the necessary information for the correct use of the instrument as well as a precise idea of its versatili...

Page 3: ...EAR PANEL 1 BNC connector for pH or ORP pH 21 only electrode 2 Temperature probe socket for use with the HI 7662 3 Power supply input Plug the 12 Vdc adapter into the DC socket then into the mains Not...

Page 4: ...nking The LCD will show the ORP mV value TAKING pH MEASUREMENTS Make sure that the instrument has been calibrated for pH before taking measurements see pH Calibration section for details Connect the s...

Page 5: ...H 10 01 buffer solution The meter automatically recognizes the buffer and displays the correct value After measurements turn the meter off by pressing the ON OFF key disconnect the power adapter from...

Page 6: ...roken insulation on the cable or cracks on the electrode stem or bulb If any scratches or cracks are present replace the electrode Rinse off any salt deposits with water Connectors must be perfectly c...

Page 7: ...ectrodes HI 8071 3 5M KCl AgCl electrolyte 4 x 30 mL for single junction electrodes FDA approved bottle HI 7072 1M KNO3 electrolyte 4 x 30 mL CLEANING PROCEDURE General Soak in HI7061 General Cleaning...

Page 8: ...s HI 8082 3 5M KCl electrolyte 4 x 30 mL for double junction electrodes FDA approved bottle ORP SOLUTIONS HI 7091M Reducing pretreatment solution 230 mL bottle HI 7091L Reducing pretreatment solution...

Page 9: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Reviews: