background image

Preliminary Examination

Remove the instrument and accessories from the packaging and 
examine it carefully. For further assistance, please contact your local 
Hanna Instruments office or email us at [email protected].
Each instrument is delivered in a rugged case and is supplied with:
• 

HI1230B

 pH electrode

• 

HI7662

 temperature probe

•  pH 4.01 buffer solution (1 sachet)

•  pH 7.01 buffer solution (1 sachet)

•  Cleaning solution (2 sachets)

•  9V Alkaline battery (1 pc.)

•  Calibration screwdriver

•  Instrument quality certificate
•  Instruction manual

Note: 

Save all packing material until you are sure that the meter 

works correctly. Any damaged or defective item must be returned in 

its original packing material with the supplied accessories.

General Description & Intended Use

HI83141-1

 is a portable, water-resistant meter designed to be paired 

with 

HI1230B

 combination pH electrode and 

HI7662

 temperature 

probe for pH, mV, and temperature measurements. Optionally, the 
meter can be paired with 

HI3131B

 refillable ORP electrode (not 

supplied with the product) for accurate ORP measurements.

Main Features

•  Two-point calibration (front panel calibration trimmers)
•  Hand-strap on the bottom for a more secure grip
•  Compact, heavy-duty, water-resistant instrument
•  Ideal for applications that require a custom calibration point

Probe Features

HI1230B

 

pH

•  Durable, polyetherimide (PEI) resin body suitable for a wide range of 

applications and chemically resistant to many aggressive chemicals.

•  Double junction reference. No silver from the electrode enters the 

sample. This design allows measurement in applications where 
silver ions in the sample are undesirable or silver precipitates on 
the junction are likely to form.

HI7662

 

Temperature

•  Build-in NTC Thermistor for automatic temp. compensation (ATC)
•  Stainless steel
•  RCA connector

Specifications

Meter

pH

Range *

0.00 to 14.00 pH

Resolution

0.01 pH

Accuracy

±0.01 pH (@25 °C / 77 °F)

Temperature 

compensation Automatic, from 0 to 100 °C (32 to 212 °F)
Calibration

Manual, 2-point (trimmers)

Offset: ±1.00 pH; Slope: 80 to 110%

mV

Range *

±1999 mV

Resolution

1 mV

Accuracy 

±1 mV (@25 °C / 77 °F)

Temperature

Range *

0.0 to 100.0 °C (32.0 to 212.0 °F)

Resolution

0.1 °C / 0.1 °F

Accuracy

±0.4 °C / ±0.8 °F (excluding probe error)

Probes (included)

HI1230B

 pH electrode; 

HI7662

 temperature

Battery type

1 x 9V Alkaline

Battery life

Approximately 150 hours of continuous use

Environment

0 to 50 °C (32 to 122 °F); 100% RH

Auto shut-off

After 8 minutes of non-use

Dimensions

145 x 80 x 36 mm (5.7 x 3.1 x 1.4’’)

Weight

230 g (8.1 oz.)

* The range may be limited by the probe’s limits.

HI1230B pH Probe

Reference

double, Ag/AgCl

Junction

ceramic, single

Electrolyte

gel

Recommended 

Operating Temp.

–5 to 70 °C (23 to 158 °F)

Max. Pressure

2 bar

Range

0 to 12 pH

Tip / Shape

spheric, Ø 7.5 mm (0.29”)

Glass type

LT (low temperature)

Body material

PEI

Dimensions

Length: 163.5 mm (6.43“)

Shaft length: 120 mm (4.72“)

Shaft diameter: 12 mm (0.47“)

Cable length

1 m (3.3’); coaxial

HI7662 Temperature Probe

Dimensions

Total length: 173.5 mm (6.8”)

Active part: 105 mm (4.1”); Ø 3.6 mm (1.4”)

Cable length

1 m (3.3’)

Functional Description & LCD Display

HI83141-1

HI83141-1

pH Meter

mV

On/Off

pH

TEMP

pH   0.00 to 14.00

mV  ±1999

°C    0.0 to 100.0 

SLOPE

OFFSET

CAL

1

14

13

12

2
3
4

11

5
6

7

10

8

9

1.  BNC connector (pH or ORP electrode)

2.  Temperature compensation status (ATC)

3.  Battery indicator (BATT)

4.  Temperature unit (°C or °F)

5. 

pH

 key – pH range selection

6. 

mV

 key – mV (ORP) range selection

7. 

TEMP

 key –temperature measurement & unit selection

8. 

OFFSET

 trimmer – adjusts offset (pH calibration)

9. 

SLOPE

 trimmer – adjusts slope (pH calibration)

10. 

CAL

 key – enters pH calibration

11. 

On/Off 

key

12. LCD display

13. Measurement mode (mV or pH)
14. RCA connector (temperature probe)

General Operations

Battery Replacement

1.  Turn off the instrument.

2.  Remove the three screws on the 

back of the instrument to open 

the battery compartment.

3. Remove the old battery. Insert 

one new 9 V Alkaline battery 

in the battery compartment 

while paying attention to the 

correct polarity.

4.  Close the battery compartment 

using the three screws.

Note: 

If the “BATT” tag is displayed 

blinking, battery level is too low and the 

battery needs to be replaced.

Connecting the pH Electrode & Temperature Probe

Align the 

HI1230B

 BNC connector with the socket and plug it in. 

Rotate the locking shell until it clicks in place.
Plug the 

HI7662

 probe RCA connector into the designated socket.

Turning the Meter On

Press the 

On/Off

 key to turn the meter on. 

Initialization screen briefly displays all the 
LCD segments followed by the battery status, 
alerting user to the remaining battery life. 

Sensor Preparation & Conditioning

1.  Remove the protective cap.

2.  If the protective cap does not contain any liquid, pour 

HI70300

 

Storage solution into the cap.

3.  Place it back on the sensor and soak for at least 30 minutes 

before use.

4.  Rinse with tap water prior to Calibration or Measurement.

pH Calibration

For high accuracy, frequent calibrations are recommended.
Additionally, the pH range should be recalibrated:

•  whenever the pH electrode is replaced

•  at least once a month
•  after testing aggressive chemicals

Preparation

One- or two-point calibration can be performed using one of the 
following standard buffer solutions: pH 4.01, 6.86 (NIST), 7.01, 
9.18 (NIST), or pH 10.01.
When a two-point calibration is required, use pH 7.01 or pH 6.86 
buffer as first calibration point.
Use pH 7.01 (

HI7007

) (or pH 6.86 NIST equivalent) for neutral 

samples, pH 4.01 (

HI7004

) for acidic samples, pH 10.01 (

HI7010

(or pH 9.18 NIST equivalent) for alkaline samples.
For best results, use a rinse beaker and a separate calibration beaker 
for each buffer. Discard rinsing buffers after use.

Procedure

1.  Connect the probes and turn the meter on.
2.  Remove the protective cap and rinse the tip with first buffer being 

used for calibration (e.g. pH 7.01).

3. Place the tip of the electrode 4 cm (1 

1/2

”) 

into correct buffer. Allow a few minutes 
for the probe and buffer to stabilize/to 
reach thermal equilibrium.

4.  Press the 

CAL

 key. The calibration range 

is automatically recognized and the 
offset on-screen tutorial messages are displayed.

5.  Press the 

TEMP

 key to read the buffer 

temperature. Take a note of the displayed value.

6.  Press the 

pH

 key to take the pH reading. 

Stir gently.

 

TEMP

pH

Dear Customer, 

Thank you for choosing a Hanna Instruments

®

 product. Please 

read this instruction manual carefully before using this instrument. 

For more information about Hanna Instruments and our products, 

visit www.hannainst.com or e-mail us at [email protected].

For technical support, contact your local Hanna Instruments office 

or e-mail us at [email protected].

Summary of Contents for HI83141-1

Page 1: ...HI83141 1 Portable pH mV Temperature Meter INSTRUCTION MANUAL ...

Page 2: ...I3131B refillable ORP electrode not supplied with the product for accurate ORP measurements Main Features Two point calibration front panel calibration trimmers Hand strap on the bottom for a more secure grip Compact heavy duty water resistant instrument Ideal for applications that require a custom calibration point Probe Features HI1230B pH Durable polyetherimide PEI resinbodysuitableforawiderang...

Page 3: ...tely 150 hours of continuous use Environment 0 to 50 C 32 to 122 F 100 RH Auto shut off After 8 minutes of non use Dimensions 145 x 80 x 36 mm 5 7 x 3 1 x 1 4 Weight 230 g 8 1 oz The range may be limited by the probe s limits HI1230B pH Probe Reference double Ag AgCl Junction ceramic single Electrolyte gel Recommended Operating Temp 5 to 70 C 23 to 158 F Max Pressure 2 bar Range 0 to 12 pH Tip Sha...

Page 4: ...set pH calibration 9 SLOPE trimmer adjusts slope pH calibration 10 CAL key enters pH calibration 11 On Off key 12 LCD display 13 Measurement mode mV or pH 14 RCA connector temperature probe General Operations Battery Replacement 1 Turn off the instrument 2 Remove the three screws on the back of the instrument to open the battery compartment 3 Remove the old battery Insert one new 9 V Alkaline batt...

Page 5: ...s Preparation One or two point calibration can be performed using one of the following standard buffer solutions pH 4 01 6 86 NIST 7 01 9 18 NIST or pH 10 01 When a two point calibration is required use pH 7 01 or pH 6 86 buffer as first calibration point Use pH 7 01 HI7007 or pH 6 86 NIST equivalent for neutral samples pH 4 01 HI7004 for acidic samples pH 10 01 HI7010 or pH 9 18 NIST equivalent f...

Page 6: ...meter on 2 Afterinitialization batterystatusisdisplayed Replace the battery if BATT tag is displayed blinking 3 Connect the probe pH or ORP to the meter 4 For accurate pH measurements calibrate the meter for pH before use 5 Place the tip of the electrodes 4 cm 1 1 2 into the sample 6 Press the designated key to select corresponding measurement mode i e pH for pH measurements mV for mV or ORP measu...

Page 7: ...ering info Product description HI1230B Plastic body double junction gel combination pH electrode with BNC connector and 1 m 3 28 ft cable HI3131B Glass body refillable combination platinum ORP electrode with BNC connector and 1 m 3 28 ft cable HI7662 Temperature probe with 1 m 3 28 ft cable HI7004M pH 4 01 buffer solution 230 mL HI7006M pH 6 86 buffer solution 250 ml HI7007M pH 7 01 buffer solutio...

Page 8: ...onments Warranty HI83141 1 is warranted for a period of two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions Electrodes and probes are warranted for a period of six months This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not cov...

Reviews: