Preliminary Examination
Remove the instrument and accessories from the packaging and
examine it carefully. For further assistance, please contact your local
Hanna Instruments office or email us at [email protected].
Each instrument is delivered in a rugged case and is supplied with:
•
HI1230B
pH electrode
•
HI7662
temperature probe
• pH 4.01 buffer solution (1 sachet)
• pH 7.01 buffer solution (1 sachet)
• Cleaning solution (2 sachets)
• 9V Alkaline battery (1 pc.)
• Calibration screwdriver
• Instrument quality certificate
• Instruction manual
Note:
Save all packing material until you are sure that the meter
works correctly. Any damaged or defective item must be returned in
its original packing material with the supplied accessories.
General Description & Intended Use
HI83141-1
is a portable, water-resistant meter designed to be paired
with
HI1230B
combination pH electrode and
HI7662
temperature
probe for pH, mV, and temperature measurements. Optionally, the
meter can be paired with
HI3131B
refillable ORP electrode (not
supplied with the product) for accurate ORP measurements.
Main Features
• Two-point calibration (front panel calibration trimmers)
• Hand-strap on the bottom for a more secure grip
• Compact, heavy-duty, water-resistant instrument
• Ideal for applications that require a custom calibration point
Probe Features
HI1230B
pH
• Durable, polyetherimide (PEI) resin body suitable for a wide range of
applications and chemically resistant to many aggressive chemicals.
• Double junction reference. No silver from the electrode enters the
sample. This design allows measurement in applications where
silver ions in the sample are undesirable or silver precipitates on
the junction are likely to form.
HI7662
Temperature
• Build-in NTC Thermistor for automatic temp. compensation (ATC)
• Stainless steel
• RCA connector
Specifications
Meter
pH
Range *
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy
±0.01 pH (@25 °C / 77 °F)
Temperature
compensation Automatic, from 0 to 100 °C (32 to 212 °F)
Calibration
Manual, 2-point (trimmers)
Offset: ±1.00 pH; Slope: 80 to 110%
mV
Range *
±1999 mV
Resolution
1 mV
Accuracy
±1 mV (@25 °C / 77 °F)
Temperature
Range *
0.0 to 100.0 °C (32.0 to 212.0 °F)
Resolution
0.1 °C / 0.1 °F
Accuracy
±0.4 °C / ±0.8 °F (excluding probe error)
Probes (included)
HI1230B
pH electrode;
HI7662
temperature
Battery type
1 x 9V Alkaline
Battery life
Approximately 150 hours of continuous use
Environment
0 to 50 °C (32 to 122 °F); 100% RH
Auto shut-off
After 8 minutes of non-use
Dimensions
145 x 80 x 36 mm (5.7 x 3.1 x 1.4’’)
Weight
230 g (8.1 oz.)
* The range may be limited by the probe’s limits.
HI1230B pH Probe
Reference
double, Ag/AgCl
Junction
ceramic, single
Electrolyte
gel
Recommended
Operating Temp.
–5 to 70 °C (23 to 158 °F)
Max. Pressure
2 bar
Range
0 to 12 pH
Tip / Shape
spheric, Ø 7.5 mm (0.29”)
Glass type
LT (low temperature)
Body material
PEI
Dimensions
Length: 163.5 mm (6.43“)
Shaft length: 120 mm (4.72“)
Shaft diameter: 12 mm (0.47“)
Cable length
1 m (3.3’); coaxial
HI7662 Temperature Probe
Dimensions
Total length: 173.5 mm (6.8”)
Active part: 105 mm (4.1”); Ø 3.6 mm (1.4”)
Cable length
1 m (3.3’)
Functional Description & LCD Display
HI83141-1
HI83141-1
pH Meter
mV
On/Off
pH
TEMP
pH 0.00 to 14.00
mV ±1999
°C 0.0 to 100.0
SLOPE
OFFSET
CAL
1
14
13
12
2
3
4
11
5
6
7
10
8
9
1. BNC connector (pH or ORP electrode)
2. Temperature compensation status (ATC)
3. Battery indicator (BATT)
4. Temperature unit (°C or °F)
5.
pH
key – pH range selection
6.
mV
key – mV (ORP) range selection
7.
TEMP
key –temperature measurement & unit selection
8.
OFFSET
trimmer – adjusts offset (pH calibration)
9.
SLOPE
trimmer – adjusts slope (pH calibration)
10.
CAL
key – enters pH calibration
11.
On/Off
key
12. LCD display
13. Measurement mode (mV or pH)
14. RCA connector (temperature probe)
General Operations
Battery Replacement
1. Turn off the instrument.
2. Remove the three screws on the
back of the instrument to open
the battery compartment.
3. Remove the old battery. Insert
one new 9 V Alkaline battery
in the battery compartment
while paying attention to the
correct polarity.
4. Close the battery compartment
using the three screws.
Note:
If the “BATT” tag is displayed
blinking, battery level is too low and the
battery needs to be replaced.
Connecting the pH Electrode & Temperature Probe
Align the
HI1230B
BNC connector with the socket and plug it in.
Rotate the locking shell until it clicks in place.
Plug the
HI7662
probe RCA connector into the designated socket.
Turning the Meter On
Press the
On/Off
key to turn the meter on.
Initialization screen briefly displays all the
LCD segments followed by the battery status,
alerting user to the remaining battery life.
Sensor Preparation & Conditioning
1. Remove the protective cap.
2. If the protective cap does not contain any liquid, pour
HI70300
Storage solution into the cap.
3. Place it back on the sensor and soak for at least 30 minutes
before use.
4. Rinse with tap water prior to Calibration or Measurement.
pH Calibration
For high accuracy, frequent calibrations are recommended.
Additionally, the pH range should be recalibrated:
• whenever the pH electrode is replaced
• at least once a month
• after testing aggressive chemicals
Preparation
One- or two-point calibration can be performed using one of the
following standard buffer solutions: pH 4.01, 6.86 (NIST), 7.01,
9.18 (NIST), or pH 10.01.
When a two-point calibration is required, use pH 7.01 or pH 6.86
buffer as first calibration point.
Use pH 7.01 (
HI7007
) (or pH 6.86 NIST equivalent) for neutral
samples, pH 4.01 (
HI7004
) for acidic samples, pH 10.01 (
HI7010
)
(or pH 9.18 NIST equivalent) for alkaline samples.
For best results, use a rinse beaker and a separate calibration beaker
for each buffer. Discard rinsing buffers after use.
Procedure
1. Connect the probes and turn the meter on.
2. Remove the protective cap and rinse the tip with first buffer being
used for calibration (e.g. pH 7.01).
3. Place the tip of the electrode 4 cm (1
1/2
”)
into correct buffer. Allow a few minutes
for the probe and buffer to stabilize/to
reach thermal equilibrium.
4. Press the
CAL
key. The calibration range
is automatically recognized and the
offset on-screen tutorial messages are displayed.
5. Press the
TEMP
key to read the buffer
temperature. Take a note of the displayed value.
6. Press the
pH
key to take the pH reading.
Stir gently.
TEMP
pH
Dear Customer,
Thank you for choosing a Hanna Instruments
®
product. Please
read this instruction manual carefully before using this instrument.
For more information about Hanna Instruments and our products,
visit www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments office
or e-mail us at [email protected].