HI83141-1
Portable
pH / mV / Temperature
Meter
INSTRUCTION MANUAL
7. Wait a few minutes and use the calibration screwdriver to adjust the
OFFSET trimmer until it displays the
pH value at the previously noted
temperature. See pH Buffer Values
at Various Temperatures table.
8. Press the
pH
key.
9. Rinse with second buffer being used for calibration (e.g. pH 4.01
or pH 10.01). Place the tip of the electrode 4 cm (1
1/2
”) into the
second buffer. Stir gently.
10. Press
TEMP
key to read buffer temperature. Take a note of the
displayed value.
11. Press
pH
key to take a pH reading.
12. Wait a couple of minutes and
adjust the SLOPE trimmer until
it displays the pH value at the
previously noted temperature.
See pH Buffer Values at Various
Temperatures table.
13. Press
CAL
key. The pH calibration is now complete.
Measurement
1. Turn the meter on.
2. After initialization, battery status is displayed.
Replace the battery if BATT tag is displayed
blinking.
3. Connect the probe (pH or ORP) to the meter.
4. For accurate pH measurements, calibrate the
meter for pH before use.
5. Place the tip of the electrodes 4 cm (1
1/2
”) into the sample.
6. Press the designated key to select corresponding measurement
mode (i.e.
pH
for pH measurements,
mV
for mV or ORP
measurements, and
TEMP
for temperature measurements).
pH
mV
TEMP
7. Stir briefly and wait a few minutes for the measurement to
stabilize. Measured value and unit tag is displayed for the
selected parameter.
Note:
If measurements are taken in different samples successively,
rinse the probes between measurements to avoid cross
contamination. After cleaning, rinse the probes with some of the
samples to be measured next.
Warnings
If measured value is outside the parameter limit of the probe, the
maximum or minimum value is displayed blinking.
OFFSET
SLOPE
4 cm (1
½
”)
pH Buffer Values at Various Temperatures
Temp
pH Values
°C
°F
4.01
7.01
10.01
0
32
4.01
7.13
10.32
5
41
4.00
7.10
10.24
10
50
4.00
7.07
10.18
15
59
4.00
7.04
10.12
20
68
4.00
7.03
10.06
25
77
4.01
7.01
10.01
30
86
4.02
7.00
9.96
35
95
4.03
6.99
9.92
40
104
4.04
6.98
9.88
45
113
4.05
6.98
9.85
50
122
4.06
6.98
9.82
55
131
4.07
6.98
9.79
60
140
4.09
6.98
9.77
65
149
4.11
6.99
9.76
70
158
4.12
6.99
9.75
75
167
4.14
7.00
9.74
80
176
4.16
7.01
9.73
85
185
4.17
7.02
9.74
90
194
4.19
7.03
9.75
95
203
4.20
7.04
9.76
Accessories
Ordering info.Product description
HI1230B
Plastic body, double-junction, gel, combination pH
electrode with BNC connector and 1 m (3.28 ft) cable
HI3131B
Glass-body, refillable, combination platinum ORP
electrode with BNC connector and 1 m (3.28 ft) cable
HI7662
Temperature probe with 1 m (3.28 ft) cable
HI7004M
pH 4.01 buffer solution, 230 mL
HI7006M
pH 6.86 buffer solution, 250 ml
HI7007M
pH 7.01 buffer solution, 230 mL
HI7009M
pH 9.18 buffer solution, 250 ml
HI7010M
pH 10.01 buffer solution, 230 mL
HI70300M
Storage solution, 230 mL
HI7061M
General cleaning solution, 230 mL
HI7091L
Reducing pretreatment solution,
500 mL + 14 g (set)
HI7092M
Oxidizing pretreatment solution, 250 mL
HI731326
Calibration screwdriver (20 pcs.)
HI76405
Electrode holder
Certification
All Hanna
®
instruments conform to the
CE European Directives.
Disposal of Electrical & Electronic Equipment.
The product should
not be treated as household waste. Instead, hand it over to the
appropriate collection point for the recycling of electrical and electronic
equipment, which will conserve natural resources.
Disposal of waste batteries.
This product contains batteries, do not
dispose of them with other household waste. Hand them over to the
appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential
negative consequences for the environment and human health. For
more information, contact your city, your local household waste
disposal service, or the place of purchase.
Recommendations for Users
Before using this meter, make sure that it is entirely suitable for
your specific application and for the environment in which it is used.
Any variation introduced by the user to the supplied equipment may
degrade the meter’s performance. For your and the meter’s safety do
not use or store the meter in hazardous environments.
Warranty
HI83141-1
is warranted for a period of two years against defects
in workmanship and materials when used for its intended purpose
and maintained according to instructions. Electrodes and probes
are warranted for a period of six months. This warranty is limited
to repair or replacement free of charge. Damage due to accidents,
misuse, tampering, or lack of prescribed maintenance is not covered.
If service is required, contact your local Hanna Instruments
®
office.
If under warranty, report the model number, date of purchase, serial
number and the nature of the problem. If the repair is not covered
by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments office, first obtain
a Returned Goods Authorization (RGA) number from the Technical
Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.
Hanna Instruments reserves the right to modify the design,
construction, or appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited without
the written consent of the copyright owner, Hanna Instruments Inc.,
Woonsocket, Rhode Island, 02895, USA.
IST83141-1 01/22