Autohold with Manual logging
1. Select
Direct/Autohold
in the Reading Mode (
Reading
tab).
2. Select
Manual
under Logging Type (
Logging
tab).
The meter will find a measurement to hold once the meter meets its stability criteria.
Autohold
symbol is displayed.
3. Tap displayed
M
symbol to take a manual log of the held measurement.
Hold values will be denoted with “H” in the log notes.
Note:
M
can be tapped either before or after the meter has found its hold value.
4. Tap
DO
to restore direct measurements.
12.3. AUTOHOLD LOGGING
1. From Measurement screen, tap (Measurement Menu).
2. Tap
Reading
tab to select stability criteria (Accurate, Medium, or Fast).
Note:
Autohold logging uses this criteria for logging. Setting this will affect when data is recorded.
3. Tap to select
Direct/Autohold
reading mode.
4. Tap
Logging
tab and select
Autohold
logging type.
5. Create a file name or have a file previously set up.
6. Option to define sample ID (name and prefix)
• Tap
Sample ID Prefix
field to start editing.
• Use the on‑screen keypad to enter log file name.
• Tap
to save new name.
• Then, scroll to select Prefix value.
7. Tap (
Home
key) to reenter Measurement screen.
8. Move probe to new sample.
9. Tap to enable autohold.
10. Tap
D
to enable autohold logging. Autohold starts flashing until autohold is reached.
11. Once logged, tap
DO
to disable autohold.