55
Logging
Three logging types are available: Automatic, Manual, and Autohold.
Automatic logging
• Readings are logged ( ) at predefined period intervals.
Sampling interval options range from 1 second to 180 minutes.
• Records are continuously added to it until the session stops.
• For each automatic logging session, a new file is created.
• A complete set of GLP information including date, time, mode selection, temperature reading, and calibration
information is stored with each log. User and log information e.g. company, address, sample details may
also be included on the .CSV file.
Manual logging
• Readings are logged each time
M
is tapped.
• All manual readings are stored in a single file (i.e. records made on different days share the same file).
Autohold logging
• Readings are logged each time
D
is tapped and configured stability criteria is reached.
• All Autohold readings are stored in a single log (i.e. records made on different days are logged in the same file).
Note:
Stability criteria can be set to Fast, Medium, or Accurate for all logging types.
12.1. AUTOMATIC LOGGING
1. From Measurement screen, tap (Measurement Menu).
2. Tap
Logging
tab and select
Automatic
logging type.
3. Scroll to select Sampling Period.
4. Tap (
Home
key) to reenter Measurement screen.
5. From Measurement screen, tap symbol to start / stop.