
39
Measurement & Probe Setup Menu
9. MEASUREMENT & PROBE SETUP MENU
From the Measurement screen, tap to access system and measurement configuration tabs.
Alternatively, tap (
Menu
key) then .
ATC
°C
25.0
[M]
mmHg
760.0
ppm
8.74
DO
Hanna_01
Stable
Admin
1%
12:40:12
10/01/2023
9.1. CALIBRATION
Options: Last Calibration, Standard Entry Type, Calibration Reminder, User Pressure Calibration
Last Calibration:
calibrate or clear a previous calibration
Standard Entry Type
• Automatic
: meter selects the closest calibration standard value to that of the sample being measured
• Manual
: user selects the standard to be used for calibration
Calibration Reminder
•
Disabled
• Daily‑
calibration interval (hours and minutes)
• Periodic
‑calibration interval (days, hours, minutes)
“Calibrate probe” message is displayed if calibration is not done within set interval.
User Pressure Calibration:
option available with Pressure Source set to Automatic (under
Reading
tab)