Measurement & Probe Setup Menu
42
Temperature Unit: Celsius, Fahrenheit, Kelvin
degrees
User Temperature Calibration: Calibrate
or
Clear
(with Automatic Temperature Source selected only)
New temperature calibration procedure
1. Tap
Calibrate
.
2. Place the probe and a reference thermometer into a stirred container of water.
Allow for the reading to stabilize.
3. If the displayed value is different than the reference thermometer reading, tap
Calibrated Temperature
and use the on‑screen keypad to adjust the value to that given by the reference thermometer.
4. Tap
Save
to confirm and save the calibration data.
9.4. VIEW
Options: Basic, Simple GLP, Full GLP, Graph, Table
• Select preferred display configuration from View Type window.