19
• After the third calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the
CFM
functional key is displayed.
• Press
CFM
to confirm calibration.
• The calibrated value and the fifth expected buffer will be displayed.
• After the fourth calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the fifth buffer solution and
stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the
CFM
functional key is displayed.