32
LAST RELATIVE mV CALIBRATION DATA
Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press
GLP
key while in Relative mV
measurement mode.
The instrument will display the Relative mV GLP information: calibration date,
time and offset.
LAST ISE CALIBRATION DATA
Last ISE calibration data is stored automatically after a successful calibration.
To view the ISE calibration data, press
GLP
while in ISE measurement mode.
The instrument will display the ISE calibration information: calibration date,
time, slope, calibration status and electrode type.
Notes:
• Press
GLP
or
ESC
at any moment and the instrument will return
to measurement mode.
• If calibration has not been performed, the instrument displays
“No user calibration”
message.
• The calibration standards from previous calibration are displayed
in video inverse mode.