background image

16

• Remove the cuvette.

• Place the 

CAL CHECK™

 Standard

HI 96702-11

 Cuvette B into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press 

CAL CHECK

 and the lamp, cuvette

and detector icons together with “

CAL

CHECK

” will appear on the display,

depending on the measurement phase.

• At the end of the measurement the display

will show the validation standard value.

The reading should be within specifications as reported in the 

CAL CHECK™

 Standard

Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.

Summary of Contents for HI 96702

Page 1: ...1 Instruction Manual HI 96702 Copper High Range ISM w w w h a n n a i n s t c o m...

Page 2: ...RY MANAGEMENT 21 BATTERY REPLACEMENT 22 ACCESSORIES 22 WARRANTY 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use...

Page 3: ...ch HI 96702 Ion Selective Meter is supplied complete with Two Sample Cuvettes and Caps CAL CHECK standard cuvettes 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruct...

Page 4: ...t by a transparent cup This makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrument...

Page 5: ...t mode after 1 hour of non use in calibration mode with last reading reminder Dimensions 192 x 104 x 69 mm 7 6 x 4 1 x 2 7 Weight 360 g 12 7 oz REQUIRED REAGENTS Code Description Quantity test HI93702...

Page 6: ...at wavelength c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are...

Page 7: ...coming value into the desired measuring unit and to display it on the LCD The measurement process is carried out in two phases first the meter is zeroed and then the actual measurement is performed Th...

Page 8: ...ode press to edit the date and time OPERATING MODES Measurement mode default operation mode enables both validation and measurement Calibration mode may be entered by keeping CAL CHECK pressed for thr...

Page 9: ...measurement Please check the preparation of the zero cuvette Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvette No Light The instrument cann...

Page 10: ...n over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standard Low The standard reading is less tha...

Page 11: ...Proper use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet It is important that the sample doe...

Page 12: ...the reaction times reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a me...

Page 13: ...unreacted sample up to the mark and replace the cap Place the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO CFM and the lamp c...

Page 14: ...to measurement The beeper is playing a beep at the end of countdown period Alternatively wait for 45 seconds then just press READ TIMER In both cases the lamp cuvette and detector icons will appear on...

Page 15: ...meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Place the CAL CHECK Standard HI 96702 11 Cuvette A into the cuvette holder and ensure that the...

Page 16: ...ll appear on the display depending on the measurement phase At the end of the measurement the display will show the validation standard value The reading should be within specifications as reported in...

Page 17: ...When the beeper sounds briefly and the LCD displays dashes the meter is ready Press and hold CAL CHECK for three seconds to enter calibration mode The display will show CAL during calibration procedur...

Page 18: ...e display depending on the measurement phase After measurement the instrument will show for three seconds the CAL CHECK Standard value Note If the display shows STD HIGH the standard value was too hig...

Page 19: ...When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press GLP to edit the desired day 01 31 If the key is kept pressed the day numbe...

Page 20: ...y and the instrument returns to measurement mode after three seconds It is possible to delete the calibration and restore factory calibration Press GLP to enter GLP mode Press READ TIMER to enter in t...

Page 21: ...ed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The remaining battery...

Page 22: ...urn it clockwise to close To replace the instrument s battery follow the steps ACCESORIES REAGENT SET HI93702 01 Reagentsfor100tests HI93702 03 Reagentsfor300tests OTHER ACCESORIES HI96702 11 CAL CHEC...

Page 23: ...tained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If serv...

Page 24: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst co...

Reviews: