14
• Remove the cuvette.
• Add the content of one packet of
HI 93702-0
reagent. Replace the
cap and shake gently.
• Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
• Press and hold
READ/ /TIMER
for three
seconds.
The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 45 seconds then
just press
READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the instrument
directly displays concentration in mg/L of
copper on the LCD.
INTERFERENCE
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted
between 6 and 8.