The 

HI 95771 

is an auto diagnostic portable microprocessor meter that benefits from Hanna’s

years of experience as a manufacturer of analytical instruments. It has the advanced optical
system based on a special tungsten lamp and a narrow band interference filter that allows most
accurate and repeatable readings. All instruments are factory calibrated and the electronic and

optical design minimizes the need of frequent calibration.
With the powerful CAL CHECK™ 

validation

 function you are able to validate good performance of

your instrument at any time. The validation procedure is extremely user friendly. Just use the
exclusive Hanna ready-made, Nist traceable standards to verify the performance of the instrument
and recalibrate if necessary.
All instruments are splash proof and the lamp and filter units are protected from dust or dirt by
a transparent cup. This makes the instruments fulfill field applications. Display codes aid the
user in routine operation. The meter has an auto-shut off feature that will turn off the instrument
after 10 minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same
position every time it is placed into the measurement cell. It is designed to fit a cuvet with a
larger neck making it easier to add both sample and reagents. The cuvet is made from special
optical glass to obtain best results.
The 

HI 95771

 is a dual range meter that can measure the chlorine (Cl

2

) content in water

samples in the 0 to 5.00 mg/L (ppm) range and in the 0 to 500 mg/L (ppm) range. The
methods are an adaptation of 

Standard Methods for the Examination of Water and

Wastewater, 20

th

 edition,

 4500-Cl.

The reagents are in

 

powder form and are supplied in packets. The amount of reagent is precisely

dosed to ensure the maximum repeatability.

degree Celsius
degree Fahrenheit
Liquid Crystal Display
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter

°C

:

°F

:

LCD:

mg/L

:

mL

:

ABBREVIATIONS

GENERAL DESCRIPTION

21

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and

replace the cover.

9V

SLIDE

OFF

BATTERY REPLACEMENT

FACTORY CALIBRATION RESET

It is possible to restore factory calibration:
• Enter the 

calibration mode

 by holding CAL CHECK

for three seconds.

• Hold ZERO/SET for 10 seconds. The display

will show for 2 seconds “F.CAL” and "771" or
“701” appears. The factory calibration is
automatically restored and the instrument is
ready for measurement by displaying "771" or

“701” on the LCD depending to which
parameter code has been selected.

REAGENT SETS
HI 95771-01

Reagents for 100 Chlorine Ultra High Range tests

HI 95771-03

Reagents for 300 Chlorine Ultra High Range tests

HI 93701-01

Reagents for 100 Free Chlorine tests

HI 93701-03

Reagents for 300 Free Chlorine tests

OTHER ACCESSORIES
HI 95771-11

CAL CHECK™ Standard Cuvets     (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL)

ACCESSORIES

or

Summary of Contents for HI 95771C

Page 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95771C Chlorine Ultra High Range ISM MAN95771CR3 06 06 w w w h a n n a i n s t c o m...

Page 2: ...choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional techn...

Page 3: ...o avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95771 is warranted for two year...

Page 4: ...m range and in the 0 to 500 mg L ppm range The methods are an adaptation of Standard Methods for the Examination of Water and Wastewater 20th edition 4500 Cl The reagents are in powder form and are su...

Page 5: ...Press READ TIMER and SIP will blink during measurement The instrument will show for three seconds the CAL CHECK standard value i e 100 for 771 chlorine ULTRA HIGH RANGE or 1 00 for 701 chlorine LOW R...

Page 6: ...through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known 19 Note It is possible to interrupt calibration procedure at a...

Page 7: ...thus requires particular attention It is important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use...

Page 8: ...andards otherwiseerroneousresultswillbeobtained For accurate Validation and Calibration results please perform test at room temperature 18 to 25 C 64 5 to 77 0 F VALIDATION PROCEDURE Hold READ TIMER f...

Page 9: ...dealer or the nearest Hanna Customer Service Center GUIDE TO DISPLAY CODES 16 MEASUREMENT PROCEDURE Select the parameter code 701 for Chlorine LOW RANGE by holding ZERO SET see page 13 When the LCD d...

Page 10: ...econd at the end of the calibration procedure to indicate that the calibration data has been stored Error the concentration of the calibration solution used is not correct Repeat the calibration proce...

Page 11: ...ne Ultra High Range 771 or 5 00 for chlorine Low Range 701 indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flas...

Page 12: ...recommended to take a new zero reading for each sample and to use the same cuvet for zeroing and measurement After the reading it is important to discard immediately the sample otherwise the glass mig...

Reviews: