background image

8

1) Liquid Crystal Display
2) Cuvet Holder
3) Cuvet alignment indicator
4) ON/OFF key
5) ZERO/SET key
6) READ/TIMER key
7) CAL CHECK key

FUNCTIONAL DESCRIPTION

KEYPAD DESCRIPTION
• ON/OFF: to turn the meter on and off.
• ZERO/SET: this is a bi-functional key: press to zero the meter prior to measurement, or hold

the key for 3 seconds to select the desired parameter (“701” chlorine LOW RANGE or “771”

chlorine ULTRA HIGH RANGE).

• READ/TIMER: this is a bi-functional key, just press for making a measurement, or hold the

key for 3 seconds to start a pre-programmed countdown prior to measurement.

• CAL CHECK: this bi-functional key allows to perform validation of the instrument just pressing

the CAL CHECK key, or hold the key for 3 seconds to enter the 

calibration mode

.

OPERATING MODES

Measurement  mode:

 default operation mode, enables both 

validation 

and 

measurement.

Calibration mode: 

may be entered

 

by keeping CAL CHECK pressed for three seconds (the “CAL”

icon appears), it enables 

calibration 

of the instrument.

17

Use the validation procedure to ensure that the instrument is properly calibrated.

Warning

: do not validate or calibrate the instrument with standard solutions other than the Hanna

CAL CHECK™ Standards, otherwise erroneous results will be obtained.
For accurate Validation and Calibration results, please perform test at room temperature: 18 to 25°C
(64.5 to 77.0°F).

VALIDATION PROCEDURE

• Hold READ/TIMER for three seconds. The

display will show the countdown prior to
measurement.
Alternatively, wait for 1 minute and just
press READ/TIMER.
In both cases “SIP” will blink during

measurement.

• The instrument directly displays concentration

in mg/L of free chlorine on the Liquid
Crystal Display.

or

INTERFERENCES

• Alkalinity:

• Acidity:

• Hardness:

• Bromine (Br

2

):

• Iodine (I

2

):

above 1,000 mg/L CaCO

if present as bicarbonate (HCO

3

-

, sample pH < 8.3);

above 250 mg/L CaCO

if present as carbonate (CO

3

2-

, sample pH > 9.0).

In both cases, it will not reliably develop the full amount of color or it may rapidly
fade (negative error). To resolve this, neutralize the sample with diluted HCl.
above 150 mg/L CaCO

3

. May not reliably develop the full amount of color or it

may rapidly fade (negative error). To resolve this, neutralize the sample with
diluted NaOH.
in case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample

for approximately 2 minutes after adding the powder reagent.
positive error.
positive error.

• Chlorine dioxide (ClO

2

):

positive error.

• Ozone (O

3

):

• Oxidized Manganese
      and Chromium:

positive error.

positive error.

Temperature correction for Validation: corrected value = read value 

×

 [1 

+

 0.002 

×

 (20 

 T(°C))]

corrected value = read value 

×

 [1 

+

 0.001 

×

 (68 

 T(°F))]

Summary of Contents for HI 95771C

Page 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95771C Chlorine Ultra High Range ISM MAN95771CR3 06 06 w w w h a n n a i n s t c o m...

Page 2: ...choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional techn...

Page 3: ...o avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95771 is warranted for two year...

Page 4: ...m range and in the 0 to 500 mg L ppm range The methods are an adaptation of Standard Methods for the Examination of Water and Wastewater 20th edition 4500 Cl The reagents are in powder form and are su...

Page 5: ...Press READ TIMER and SIP will blink during measurement The instrument will show for three seconds the CAL CHECK standard value i e 100 for 771 chlorine ULTRA HIGH RANGE or 1 00 for 701 chlorine LOW R...

Page 6: ...through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known 19 Note It is possible to interrupt calibration procedure at a...

Page 7: ...thus requires particular attention It is important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use...

Page 8: ...andards otherwiseerroneousresultswillbeobtained For accurate Validation and Calibration results please perform test at room temperature 18 to 25 C 64 5 to 77 0 F VALIDATION PROCEDURE Hold READ TIMER f...

Page 9: ...dealer or the nearest Hanna Customer Service Center GUIDE TO DISPLAY CODES 16 MEASUREMENT PROCEDURE Select the parameter code 701 for Chlorine LOW RANGE by holding ZERO SET see page 13 When the LCD d...

Page 10: ...econd at the end of the calibration procedure to indicate that the calibration data has been stored Error the concentration of the calibration solution used is not correct Repeat the calibration proce...

Page 11: ...ne Ultra High Range 771 or 5 00 for chlorine Low Range 701 indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flas...

Page 12: ...recommended to take a new zero reading for each sample and to use the same cuvet for zeroing and measurement After the reading it is important to discard immediately the sample otherwise the glass mig...

Reviews: