background image

12

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• For a correct filling of the cuvet: the liquid in the

cuvet forms a convexity on the top; the bottom of
this convexity must be at the same level of the 10

mL mark.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

(a)

(b)

(c)

• It is important that the sample does not contain any

debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 or a lint-free cloth prior to insertion.

• Shaking the cuvet can generate bubbles in the sample,

causing higher readings. To obtain accurate
measurements, remove such bubbles by swirling or
by gently tapping the vial.

13

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

OPERATIONAL GUIDE

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “771” is ready: the Chlorine ULTRA HIGH

RANGE scale is automatically selected each time the instrument is
turned on.

• To change parameter keep ZERO/SET pressed. The parameter code

will change every three seconds, until the key is released. Select the

program for “771” for chlorine ULTRA HIGH RANGE or “701” for
chlorine LOW RANGE

• After the desired parameter code appears on the display, perform

either MEASUREMENT, VALIDATION, or CALIBRATION PROCEDURE
following the instructions reported in the related chapter.

Code

Parameter

771

Chlorine Ultra High Range

701

Chlorine Low Range

or

Important Note

: chlorine ULTRA HIGH RANGE and chlorine LOW

RANGE have to be measured separately with fresh unreacted samples
if both values are requested.

Summary of Contents for HI 95771C

Page 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95771C Chlorine Ultra High Range ISM MAN95771CR3 06 06 w w w h a n n a i n s t c o m...

Page 2: ...choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional techn...

Page 3: ...o avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95771 is warranted for two year...

Page 4: ...m range and in the 0 to 500 mg L ppm range The methods are an adaptation of Standard Methods for the Examination of Water and Wastewater 20th edition 4500 Cl The reagents are in powder form and are su...

Page 5: ...Press READ TIMER and SIP will blink during measurement The instrument will show for three seconds the CAL CHECK standard value i e 100 for 771 chlorine ULTRA HIGH RANGE or 1 00 for 701 chlorine LOW R...

Page 6: ...through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known 19 Note It is possible to interrupt calibration procedure at a...

Page 7: ...thus requires particular attention It is important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use...

Page 8: ...andards otherwiseerroneousresultswillbeobtained For accurate Validation and Calibration results please perform test at room temperature 18 to 25 C 64 5 to 77 0 F VALIDATION PROCEDURE Hold READ TIMER f...

Page 9: ...dealer or the nearest Hanna Customer Service Center GUIDE TO DISPLAY CODES 16 MEASUREMENT PROCEDURE Select the parameter code 701 for Chlorine LOW RANGE by holding ZERO SET see page 13 When the LCD d...

Page 10: ...econd at the end of the calibration procedure to indicate that the calibration data has been stored Error the concentration of the calibration solution used is not correct Repeat the calibration proce...

Page 11: ...ne Ultra High Range 771 or 5 00 for chlorine Low Range 701 indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flas...

Page 12: ...recommended to take a new zero reading for each sample and to use the same cuvet for zeroing and measurement After the reading it is important to discard immediately the sample otherwise the glass mig...

Reviews: