Hanna Instruments HI 95711C Instruction Manual Download Page 7

7

A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned
and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the
cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts it

into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvet has a very important role because it is an optical element and thus requires particular
attention. It is important that both the measurement and the calibration (zeroing) cuvets are

optically identical to provide the same measurement conditions. Whenever possible use the same
cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to avoid
measurement interference due to unwanted reflection and absorption of light. It is recommended
not to touch the cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvet to prevent any contamination.

1) Liquid Crystal Display
2) Cuvet Holder
3) Cuvet alignment indicator
4) ON/OFF key
5) ZERO/SET key
6) READ/TIMER key
7) CAL CHECK key

FUNCTIONAL DESCRIPTION

18

or

• Press READ/TIMER and “SIP” will blink on the

display.

• The instrument will show for three seconds the

CAL CHECK™ standard value.
Then the date of last calibration (e.g.: “01.08”)

appears on the display, or “01.01” if the factory
calibration was selected. In both cases the
number of the month is blinking, ready for date
input.

Note

: if display will show  “ERR” the

calibration procedure failed. Verify that the right

CAL CHECK™ Standard Cuvet

 B

 is inserted,  that

both 

A

 and 

B

 cuvets are free from fingerprints or

dirt and that they are inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the desired

month number (01-12).

• When the correct month has been set, press ZERO

to confirm. Now the display will show the day
number blinking. Keep READ/TIMER pressed to
scroll to the desired day number (01-31).

Note

: it is possible to change from 

day 

to 

month

and vice versa by pressing ZERO.

• When both the day and month have been selected,

hold CAL CHECK for three seconds to store date
and calibration values. The instrument will show
for one second “Stor”, to confirm that the new
calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying the parameter
code (“701” or “711”) on the LCD.

or

Summary of Contents for HI 95711C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95711C Free Total Chlorine ISM 24 MAN95711R3 05 06 w w w h a n n a i n s t c o m...

Page 2: ...23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the met...

Page 3: ...terial until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINATION 22 Hanna publishes a wide range of catalogs and handbo...

Page 4: ...sed to ensure the maximum repeatability degree Celsius US Environmental Protection Agency degree Fahrenheit milligrams per liter mg L is equivalent to ppm part per million milliliter C EPA F mg L mL A...

Page 5: ...value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 1 00 mg Lchlorine and a repres...

Page 6: ...r factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption o...

Page 7: ...e same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON O...

Page 8: ...nking Battery icon appears when the battery voltage is getting low 17 CALIBRATION PROCEDURE Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Enter t...

Page 9: ...ODES 16 Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Place the CAL CHECK Standard Cuvet A into the holder and ensure that the notch on the cap i...

Page 10: ...reinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 15 or Hold READ TIMER for three seconds The display will show the countdown prior...

Page 11: ...er use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet a b c GENERAL TIPS FOR AN ACCURATE MEAS...

Page 12: ...it is important to discard immediately the sample otherwise the glass might become permanently stained All the reaction times reported in this manual are referred to 20 C 68 F As a general rule of thu...

Reviews: