Hanna Instruments HI 95711C Instruction Manual Download Page 10

10

The flashing digits indicate that:

1) the month can be set
2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not
correct. Repeat the 

calibration procedure

 with the right standard

solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service

Center.

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-
to-noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check
that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check if the
right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in the

measurement procedure

 for zeroing the meter.

15

or

• Hold READ/TIMER for three seconds. The display

will show the countdown prior to measurement.

Alternatively, wait for:
Free Chlorine:

Total Chlorine:

1 minute

2 minutes and 30 seconds

Then just press READ/TIMER.
In both cases “SIP” will blink during
measurement.

• The instrument directly displays concentration

in mg/L of free or total chlorine on the Liquid
Crystal Display.

or

INTERFERENCES

• Alkalinity:

• Acidity:

• Hardness:

• Bromine (Br

2

):

• Chlorine dioxide (ClO

2

):

• Iodine (I

2

):

• Oxidized Manganese

and Chromium:

• Ozone (O

3

):

above 1,000 mg/L CaCO

if present as bicarbonate (HCO

3

-

, sample pH

< 8.3); above 250 mg/L CaCO

if present as carbonate (CO

3

2-

,

sample pH > 9.0).
In both cases, it will not reliably develop the full amount of color or it
may rapidly fade (negative error). To resolve this, neutralize the

sample with diluted HCl.
above 150 mg/L CaCO

3

. May not reliably develop the full amount of

color or it may rapidly fade (negative error). To resolve this, neutralize
the sample with diluted NaOH.
in case of water with hardness greater than 500 mg/L CaCO

3

, shake

the sample for approximately 2 minutes after adding the powder
reagent.
positive error.
positive error.
positive error.
positive error.

positive error.

Summary of Contents for HI 95711C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95711C Free Total Chlorine ISM 24 MAN95711R3 05 06 w w w h a n n a i n s t c o m...

Page 2: ...23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the met...

Page 3: ...terial until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINATION 22 Hanna publishes a wide range of catalogs and handbo...

Page 4: ...sed to ensure the maximum repeatability degree Celsius US Environmental Protection Agency degree Fahrenheit milligrams per liter mg L is equivalent to ppm part per million milliliter C EPA F mg L mL A...

Page 5: ...value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 1 00 mg Lchlorine and a repres...

Page 6: ...r factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption o...

Page 7: ...e same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON O...

Page 8: ...nking Battery icon appears when the battery voltage is getting low 17 CALIBRATION PROCEDURE Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Enter t...

Page 9: ...ODES 16 Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Place the CAL CHECK Standard Cuvet A into the holder and ensure that the notch on the cap i...

Page 10: ...reinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 15 or Hold READ TIMER for three seconds The display will show the countdown prior...

Page 11: ...er use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet a b c GENERAL TIPS FOR AN ACCURATE MEAS...

Page 12: ...it is important to discard immediately the sample otherwise the glass might become permanently stained All the reaction times reported in this manual are referred to 20 C 68 F As a general rule of thu...

Reviews: