5
Range
Free Cl
2
0.00 to 5.00 mg/L
Total Cl
2
0.00 to 5.00 mg/L
Resolution
0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L
Precision
±0.02 mg/L @ 1.00 mg/L
Typical EMC Deviation
±0.01 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Light Detector
Silicon Photocell
Method
Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.
The reaction between chlorine and the DPD reagent causes a pink tint in
the sample.
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
.
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.).
REQUIRED REAGENTS
Code
Unit
Description
Quantity/test
HI 93701-0
Free Cl
2
Free Chlorine Reagent
1 packet
HI 93711-0
Total Cl
2
Total Chlorine Reagent
1 packet
Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good
accuracy, precise results can be inaccurate.
The figure explains these definitions. In a
laboratory using a standard solution of
1.00 mg/L chlorine and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.02 mg/L.
SPECIFICATIONS
PRECISION AND ACCURACY
20
REAGENT SETS
HI 93701-01
Reagents for 100 Free Chlorine tests
HI 93701-03
Reagents for 300 Free Chlorine tests
HI 93711-01
Reagents for 100 Total Chlorine tests
HI 93711-03
Reagents for 300 Total Chlorine tests
OTHER ACCESSORIES
HI 95701-11
CAL CHECK™ Standard Cuvets for Free Chlorine (1 set)
HI 95711-11
CAL CHECK™ Standard Cuvets for Total Chlorine (1 set)
HI 710009
Blue rubber boot
HI 710010
Orange rubber boot
HI 721310
9V battery (10 pcs)
HI 731318
Tissue for wiping cuvets (4 pcs)
HI 731331
Glass cuvets (4 pcs)
HI 731335
Caps for cuvets (4 pcs)
HI 93703-50
Cuvets cleaning solution (230 mL).
Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and replace
the cover.
9V
SLIDE
OFF
BATTERY REPLACEMENT
ACCESSORIES