Hanna Instruments HI 95711C Instruction Manual Download Page 5

5

Range

Free Cl

2

0.00 to 5.00 mg/L

Total Cl

2

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Precision

±0.02 mg/L @ 1.00 mg/L

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.
The reaction between chlorine and the DPD reagent causes a pink tint in
the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode

;

after 1 hour of non-use in 

calibration mode

.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS

Code

Unit

Description

Quantity/test

HI 93701-0

Free Cl

2

Free Chlorine Reagent

1 packet

HI 93711-0

Total Cl

2

Total Chlorine Reagent

1 packet

Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good

accuracy, precise results can be inaccurate.
The figure explains these definitions. In a
laboratory using a standard solution of
1.00 mg/L chlorine and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.02 mg/L.

SPECIFICATIONS

PRECISION AND ACCURACY

20

REAGENT SETS
HI 93701-01

Reagents for 100 Free Chlorine tests

HI 93701-03

Reagents for 300 Free Chlorine tests

HI 93711-01

Reagents for 100 Total Chlorine tests

HI 93711-03

Reagents for 300 Total Chlorine tests

OTHER ACCESSORIES
HI 95701-11

CAL CHECK™ Standard Cuvets     for Free Chlorine (1 set)

HI 95711-11

CAL CHECK™ Standard Cuvets     for Total Chlorine (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL).

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and replace
the cover.

9V

SLIDE

OFF

BATTERY REPLACEMENT

ACCESSORIES

Summary of Contents for HI 95711C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95711C Free Total Chlorine ISM 24 MAN95711R3 05 06 w w w h a n n a i n s t c o m...

Page 2: ...23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the met...

Page 3: ...terial until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINATION 22 Hanna publishes a wide range of catalogs and handbo...

Page 4: ...sed to ensure the maximum repeatability degree Celsius US Environmental Protection Agency degree Fahrenheit milligrams per liter mg L is equivalent to ppm part per million milliliter C EPA F mg L mL A...

Page 5: ...value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 1 00 mg Lchlorine and a repres...

Page 6: ...r factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption o...

Page 7: ...e same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON O...

Page 8: ...nking Battery icon appears when the battery voltage is getting low 17 CALIBRATION PROCEDURE Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Enter t...

Page 9: ...ODES 16 Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Place the CAL CHECK Standard Cuvet A into the holder and ensure that the notch on the cap i...

Page 10: ...reinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 15 or Hold READ TIMER for three seconds The display will show the countdown prior...

Page 11: ...er use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet a b c GENERAL TIPS FOR AN ACCURATE MEAS...

Page 12: ...it is important to discard immediately the sample otherwise the glass might become permanently stained All the reaction times reported in this manual are referred to 20 C 68 F As a general rule of thu...

Reviews: