Hanna Instruments HI 95711C Instruction Manual Download Page 12

12

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free
of fingerprints, oil or dirt. Wipe it thoroughly
with 

HI 731318

 or a lint-free cloth prior to

insertion.

• Shaking the cuvet can generate bubbles in the

sample, causing higher readings. To obtain
accurate measurements, remove such bubbles by
swirling or by gently tapping the vial.

• Do not let the reacted sample stand too long

after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row,

but it is recommended to take a new zero reading
for each sample and to use the same cuvet for
zeroing and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might

become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of
thumb, they should be doubled at 10°C (50°F)
and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the 

validation procedure

to be sure that the instrument is properly calibrated.

If necessary, calibrate the instrument.

13

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “701” or “711” it is

ready. The display code that appears is the last
selected parameter.

• Keep ZERO/SET pressed to change parameter.

The parameter code will change every three
seconds, until the key is released. Select the
program for free chlorine “701” or total chlorine
“711”.

• After the desired parameter code appears on

the display, perform either MEASUREMENT,
VALIDATION, or CALIBRATION PROCEDURE
following the instructions reported in the related
chapter.

Code

Parameter

701

Free Chlorine

711

Total Chlorine

or

OPERATIONAL GUIDE

Summary of Contents for HI 95711C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95711C Free Total Chlorine ISM 24 MAN95711R3 05 06 w w w h a n n a i n s t c o m...

Page 2: ...23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the met...

Page 3: ...terial until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINATION 22 Hanna publishes a wide range of catalogs and handbo...

Page 4: ...sed to ensure the maximum repeatability degree Celsius US Environmental Protection Agency degree Fahrenheit milligrams per liter mg L is equivalent to ppm part per million milliliter C EPA F mg L mL A...

Page 5: ...value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 1 00 mg Lchlorine and a repres...

Page 6: ...r factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption o...

Page 7: ...e same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON O...

Page 8: ...nking Battery icon appears when the battery voltage is getting low 17 CALIBRATION PROCEDURE Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Enter t...

Page 9: ...ODES 16 Select the parameter code for free chlorine 701 ot total chlorine 711 by holding ZERO SET see pag 13 Place the CAL CHECK Standard Cuvet A into the holder and ensure that the notch on the cap i...

Page 10: ...reinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 15 or Hold READ TIMER for three seconds The display will show the countdown prior...

Page 11: ...er use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet a b c GENERAL TIPS FOR AN ACCURATE MEAS...

Page 12: ...it is important to discard immediately the sample otherwise the glass might become permanently stained All the reaction times reported in this manual are referred to 20 C 68 F As a general rule of thu...

Reviews: