A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned
and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the

cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts it

into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual

measurement is performed.
The cuvet has a very important role because it is an optical element and thus requires particular

attention. It is important that both the measurement and the calibration (zeroing) cuvets are
optically identical to provide the same measurement conditions. Whenever possible use the same

cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to avoid
measurement interference due to unwanted reflection and absorption of light. It is recommended
not to touch the cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvet to prevent any contamination.

FUNCTIONAL DESCRIPTION

1) Liquid Crystal Display (LCD)
2) Cuvet Holder
3) Cuvet alignment indicator
4) ON/OFF key
5) ZERO key
6) READ/TIMER key
7) CAL CHECK key

FUNCTIONAL DESCRIPTION

14

• Replace the cuvet into the holder and ensure

that the notch on the cap is positioned securely

into the groove.

• Hold READ/TIMER 

for three seconds.

 The

display will show the countdown prior to
measurement.
Alternatively, wait for 5 minutes and just press
READ/TIMER.

In both cases “SIP” will blink during
measurement.

• The instrument directly displays concentration

in mg/L of phosphorus on the Liquid Crystal
Display.

or

INTERFERENCES

Calcium (Ca

2+

):

Chloride (Cl

-

):

Ferrous iron (Fe

2+

):

Magnesium (Mg

2+

):

Sulfide (S

2-

)

above 10,000 mg/L as CaCO

3

.

above 150,000 mg/L as Cl

-

.

above 100 mg/L as Fe

2+

.

above 40,000 mg/L as CaCO

3

.

Summary of Contents for HI 95706C

Page 1: ...1 Instruction Manual HI 95706C Phosphorus ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95706R2 06 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...ACCESSORIES HI 95706 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuv...

Page 4: ...asures the phosphorus P content in water samples in the 0 0 to 15 0 mg L ppm range The method is an adaptation from Standard Method for the Examination of Water and Wastewater The reagents are in powd...

Page 5: ...tandardsolutionof6 00 mg L phosphorus and a representative lot of reagents an operator obtained with a single instrumentastandarddeviationof0 2mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...ever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of l...

Page 8: ...ation mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCED...

Page 9: ...ration is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a...

Page 10: ...ure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zer...

Reviews: